5
C H A P T E R
Error Code
126
127
130
131
201
202
203
214
226
228
230
301
302
401
402
403
404
405
407
408
410
411
412
413
414
415
416
417
418
419
420
421
422
423
424
130
LDT-5980/ 5948
C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Troubleshoot ing
Explanation
End of data error
Invalid parameter for command
Query not supported
Command not supported
Parameter value out of range
Error in conversion of parameter type
Command is a "secure" command, but secure commands are disabled
Response is too long to output
Error in arbitrary block specification
Delimiter not found
Invalid internal response
No data found error
Not empty error
Controller output off error
Calibration failed
DSP reset failed
Calibration constant error
Invalid sensor error
Invalid mode error
DSP controller communication failure
Upper Temperature Limit caused controller output to turn off
Lower Temperature Limit caused controller output to turn off
Open sensor caused controller output to turn off
Shorted sensor caused controller outputcontroller output to turn off
Upper Current Limit caused controller output to turn off
Lower Current Limit caused controller output to turn off
Upper Voltage limit caused controller output to turn off
Lower Voltage Limit caused controller output to turn off
Open TEC Module caused controller output to turn off
Shorted TEC Module caused controller output to turn off
Sensor over resistance caused controller output to turn off
Sensor under resistance caused controller output to turn off
Plus 3V supply out-of-tolerance condition caused controller output to turn off
Plus 5V supply out-of-tolerance condition caused controller output to turn off
Plus 15V supply out-of-tolerance condition caused controller output to turn off