Thermo Scientific Orion Pro Star PH211 User Manual

Thermo Scientific Orion Pro Star PH211 User Manual

Ph/orp bench meter
Table of Contents

Advertisement

Quick Links

Operations Manual
Electrochemistry Meters
Update Meter Software to
Ensure Optimum Performance
www.thermofisher.com/orionsoftware
Orion Pro Star PH211 pH/ORP
Bench Meter User Manual
68X002725 | Revision 01 | December 2024

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Orion Pro Star PH211 and is the answer not in the manual?

Questions and answers

Summary of Contents for Thermo Scientific Orion Pro Star PH211

  • Page 1 Operations Manual Electrochemistry Meters Update Meter Software to Ensure Optimum Performance www.thermofisher.com/orionsoftware Orion Pro Star PH211 pH/ORP Bench Meter User Manual 68X002725 | Revision 01 | December 2024...
  • Page 3: Table Of Contents

    Table of Contents Chapter 1 ........................ 7 Introduction ....................7 1.1 Intended Use ..................8 1.1.1 Intended Use ....................8 1.1.2 Non-Intended Use ..................8 1.1.3 Safety ......................8 1.1.4 Power Connection ..................9 1.2 Cleaning Instructions (Meter Only) ............9 Chapter 2 ......................
  • Page 4 3.4 Data Logs ..................32 3.4.1 Filter the Data Log ..................33 3.4.2 View Individual Data Logs ................33 3.4.3 Export the Data Log ..................34 3.5 Calibration Logs ................35 3.5.1 Filter the Calibration Log ................36 3.5.2 View Individual Calibration Logs ..............37 3.5.3 Export the Calibration Log ................38 3.6 Methods ..................
  • Page 5 4.2 ORP Calibration................52 4.2.1 ORP Calibration Overview ................52 4.2.1.1 Determined ORP Standard Value ..............52 4.2.2 ORP Calibration Procedure ................52 4.3 Temperature Calibration ..............54 4.3.1 Temperature Calibration Overview ..............54 4.3.2 Temperature Calibration Procedure ..............54 Chapter 5 ......................55 Measurements ................... 55 5.1 Measure Mode Overview ..............
  • Page 6 8.2 Meter Dimensions ................73 8.3 Warranty ..................74 8.3.1 Return of Items ....................74 8.3.2 Manufacturer Information ................74 Chapter 9 ......................75 Regulatory Compliance ................75 9.1 Regional Approvals ................75 9.2 Product Safety ................75 9.3 Electromagnetic Compatibility ............75 9.3.1 FCC Statement (USA) ..................76 9.3.2 Canadian ISED IC Notice ................76 9.3.3 South Korean KC Registration ...............76...
  • Page 7: Chapter 1

    Pro Star PH211 bench meters are designed to provide accurate and reliable testing in laboratories that need a modern, easy-to-operate meter with advanced performance and reliable data reporting. Orion Pro Star PH211 bench meters offer the following measurements: • pH, mV or ORP with temperature First time meter startup is made easy with the meter setup wizard, which guides the user through important meter and measurement settings.
  • Page 8: Intended Use

    Secure a wide range of electrodes using the various electrode holder options. Neatly organize electrode cables in the built-in side channels of the stand. All Orion Pro Star PH211 meters include an electrode stand, USB flash drive, computer cable and universal 100-240V 50/60 Hz power adapter with wall plugs for US/Japan, Euro, UK/Singapore, Australia/New Zealand and China.
  • Page 9: Power Connection

    Wipe the display panel with a clean microfiber cloth, taking care to wipe lightly to avoid scratching the meter display. The meter housing can also be cleaned using a clean microfiber cloth dampened with IPA (isopropyl alcohol) for disinfection purposes. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 10: Chapter 2

    Chapter 2 Getting Started Refer to the 3.9 Information & Help section and Meter Tours to access onscreen tours that explain how to use important meter functions including meter connections, electrode stand assembly, pH electrode preparation, pH electrode storage and pH electrode cleaning. 2.1 Universal Power Supply Assembly A universal power adapter (catalog number PSTAR-PWR) with US/Japan, Euro, UK/Singapore, Australia/New Zealand and China wall plug plates is included with the meter.
  • Page 11 You are now ready to turn on the meter. CAUTION: Use of a surge protector or uninterrupted power supply (UPS) is recommended, as an unintended power surge of electricity to the meter may damage the meter and void the warranty. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 12: Electrode Stand Installation

    2.2 Electrode Stand Installation The included electrode stand (catalog number PSTAR-ARM) facilitates secure, smooth movement of electrodes into and out of solutions. A weighted base provides flexible placement of the stand around the meter. Several electrode holder and clip options accommodate a wide range of electrode types. Built-in side channels allow electrode cables to be neatly organized inside the stand.
  • Page 13 15mm diameter. e. The 20° angled holder secures gas-sensing ion selective electrodes (ISEs) including the ammonia electrode, catalog number 9512BNWP and high performance ammonia electrode, catalog number 9512HPBNWP. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 14 The RDO clip holds optical dissolved oxygen electrodes. Attach the RDO clip to the holder head by sliding the clip down on top of the tab on the front of the holder head. Use the side channels built into the electrode arm to organize the electrode cables. Press the electrode cables into the 4 grips inside the arm channels to secure the cables in place.
  • Page 15: Meter Inputs

    • Connect 1 USB flash drive with a USB-A plug to the meter for data transfer. • Connect 1 USB accessory (barcode scanner, mouse or keypad) with USB-A plug to the meter for alphanumeric character entry into meter fields including sample ID. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 16: Back Panel Meter Inputs

    2.3.3 Back Panel Meter Inputs Most meter inputs are located on the back panel of the meter. Each input has a protective cap that protects the meter from dust and water intrusion. For maximum protection, keep the protective caps securely closed unless the input is in use. Power Input: Connect the power adapter, with twist-and-lock plug, to the meter.
  • Page 17: Meter Startup Sequence

    The first time the meter is powered on, it launches the meter setup wizard. Follow the displayed steps to set critical meter and measurement settings. View detailed information on additional settings in the 3.2 Settings section. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 18: Touchscreen Functions

    2.5 Touchscreen Functions Orion Pro Star meters offer a 7” capacitive touchscreen interface that is compatible with most nitrile laboratory gloves. • Key interaction points on the screen are blue for easy navigation. • For screens with extended data, easily scroll by touching and sliding the display up and down. Optionally, a USB mouse can be connected to the meter and used to operate the touchscreen interface if direct contact with the screen is not desirable.
  • Page 19: Graph Measure Mode View

    Graph: Tap the blue Graph text to view the current measurement as a graph with temperature. • Select the Read Type to change how the graph tracks measurements over time. • Tap the Display Temperature button to show or hide temperature measurements. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 20: Basic Measure Mode View

    2.5.1.2 Basic Measure Mode View Refer to the 3.2.1 Instrument Settings section and Display settings and select Basic as the Custom Measure Layout to view a simplified version of the measure mode. The default meter setting is Advanced as the Custom Measure Layout and all user-selectable information (i.e. instrument name, sample ID, electrode ID, method) will only be shown when it is entered in the settings menus.
  • Page 21: Electrode Information

    2.6 Electrode Information The Orion Pro Star PH211 meter is compatible with a pH electrode or ORP (redox) electrode with BNC connector, ATC temperature probe with 8-pin MiniDIN connector and reference half-cell electrode with pin-tip connector. The Orion Pro Star PH211 meter is available in a meter only configuration (purchase electrode separately) or in meter kits with electrodes and solutions included for different applications.
  • Page 22: Chapter 3

    Chapter 3 Main Menu 3.1 Main Menu Tap the Main Menu icon in the measure mode to access the Main Menu. • Settings: View and customize instrument and measurement settings. • Calibrate: Perform all calibration functions, including temperature and barometer calibrations. •...
  • Page 23: Settings

    • pH Settings: View and customize measurement settings for pH, mV or ORP, including temperature, measure mode, sample ID, alarms, resolution, calibration points, stability criteria, stirrer, electrode ID, isopotential value, slope type and slope guidance. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 24: Instrument Settings

    3.2.1 Instrument Settings Tap the Instrument Settings option in the Settings list to access the instrument settings. • Read Type: Set the Read Type as Continuous, Auto-Read or Timed to define how measurements are performed, saved and exported. If Timed is selected, set the time interval from 3 seconds (0:00:03) to 2 hours (2:00:00).
  • Page 25 Custom Data Log Export: Select an option to customize the amount of information Select All Data when exporting data to LIMS software. exported from the data log. Basic: Only critical data is exported from the data log including date, time,  primary reading and temperature. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 26 Intermediate: Moderate data is exported from the data log including basic data  plus log #, username, measure mode, secondary reading (if applicable) and short calibration log summary. Advanced: Significant data is exported from the data log including intermediate  data plus read type, read time, sample ID, method, electrode ID, full calibration log summary and instrument ID.
  • Page 27: Ph Settings

    Automatic: When an ATC sensor is connected to the meter, the sensor is  shown on the list. Ambient: The meter’s built-in sensor is used to measure ambient air  temperature. While not as accurate as an ATC sensor in sample solution, it is Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 28 more accurate than not reporting temperature. Manual: When no ATC temperature sensor is connected to the meter, the meter  will default to MAN as the temperature input. Use the manual temperature input field to enter the sample temperature value, from -30.0°C to 130.0°C. •...
  • Page 29 • Slope Type: Select Segmented (point-to-point slope) or Linear Regression (single best fit slope) as the slope type to be used to calculate calibration results. o Segmented slope provides the most accurate measurements. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 30 o Linear regression uses an average slope, so it is not as accurate as using a segmented slope but may be helpful if measurements change significantly over a large range. • Slope Guidance (pH mode): Set the range for Excellent, Fair and Bad slope values. o Excellent: Default slope range is 92.0 to 102.0%.
  • Page 31: Calibrate

    3.3 Calibrate Tap the Calibrate option in the Main Menu to access the complete list of calibrations available. The Orion Pro Star PH211 pH/ORP bench meter is capable of performing: • 1 to 5 point pH calibration • 1 point ORP calibration •...
  • Page 32: Data Logs

    3.4 Data Logs Tap the Data Logs option in the Main Menu to access the data logs. The meter’s data log will store up to 10000 data logs with time and date stamps. Each data log includes measurements from when the data log was saved as well as settings, measurement information and meter details.
  • Page 33: Filter The Data Log

    Tap a log on the list to view the individual data log details. You may need to scroll down to view all the included information. Tap the blue arrows to scroll up or down through the individual data logs. Tap the Export button to export the individual data log. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 34: Export The Data Log

    3.4.3 Export the Data Log Tap the Export button to export the data log to a USB flash drive, computer or printer. • Use the Instrument Settings / Data Transfer Settings to view and modify the Export Destination, Output Format, Baud Rate (when applicable) and Custom Data Log Export settings. Secure Mode Example Open Mode Example When in Secure Mode, if an Administrator or Level 1 user account is active, that account will have the...
  • Page 35: Calibration Logs

    Tap to select a parameter to view the calibration log list for that parameter. Note: In Secure Mode, when the calibration log becomes full, the calibration log must be exported before it can be cleared. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 36: Filter The Calibration Log

    3.5.1 Filter the Calibration Log From the parameter-specific calibration log list, filter the list (Secure Mode only), view individual calibration logs and export the calibration log. Secure Mode Example Open Mode Example Tap the Filter By field to set parameters for Username (Secure Mode only) to filter the calibration log list. Chapter 3 | Main Menu...
  • Page 37: View Individual Calibration Logs

    Tap a log on the list to view the individual calibration log details. You may need to scroll down to see all included information. Tap the blue arrows to scroll up or down through the individual calibration logs. Tap the Export button to export the individual calibration log. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 38: Export The Calibration Log

    3.5.3 Export the Calibration Log Tap the Export button to export the calibration log to a USB flash drive, computer or printer. • Use the Instrument Settings / Data Transfer Settings to view and modify the Export Destination, Output Format, Baud Rate (when applicable) and Custom Calibration log Export settings. Secure Mode Example Open Mode Example When in Secure Mode, if an Administrator or Level 1 user account is active, that account will have the...
  • Page 39: Methods

    From the methods list, tap the load (triangle) icon or view (eye) icon to view the method summary information for an existing method on the list. From the method summary screen, perform one of the following actions: Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 40: Edit An Existing Method

    a. Tap the Load button to load the method to the measure mode. The meter will use the settings from the loaded method for all subsequent measurements. b. Tap the back arrow to return to the methods list without loading the method. 3.6.3 Edit an Existing Method From the methods list, tap the edit (pencil) icon to edit an existing method on the list.
  • Page 41: Secure Mode

    3 security questions for the first admin account. a. The password must contain at least 1 number, 1 uppercase letter, 1 lowercase letter, and 1 special character and must be 6-12 characters long. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 42: Create New Account (Admin Only)

    Tap the Save button. Login to the meter using the created username and password. Create new admin or user accounts, set the password expiration time, set the sleep mode in instrument settings and verify the date and time in instrument settings. 3.7.1.2 Create New Account (Admin Only) Create up to 2 additional admin accounts and up to 30 user accounts.
  • Page 43: Edit Existing Account (Admin Only)

    3.7.1.5 Set Password Expiration Time (Admin Only) Set the password expiration time to 1 month, 3 months, 6 months or 12 months. Tap the Secure Mode option in the Main Menu to access the secure mode list for admins. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 44: Deactivate Secure Mode (Admin Only)

    Tap the Secure Mode Parameters option. Expand the Password Expiration Time option and select the desired expiration period. Tap the Save button. 3.7.1.6 Deactivate Secure Mode (Admin Only) Deactivate the secure mode and revert the meter to open mode. When secure mode is deactivated, all admin and user login and logout functionality is removed, all meter access restrictions are removed, the audit trail is erased and all user accounts are deactivated.
  • Page 45: Notifications

    Error notifications are triggered when there is a critical issue that needs immediate attention that the user must address before continuing to use the meter. Example: When the pH measurement is outside the acceptable range of -2 to 20 pH. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 46: Information & Help

    3.9 Information & Help Tap the Information & Help option in the Main Menu to access the meter information, user manual, meter tours, meter tests and troubleshooting guides, setup wizard, software updates and meter reset. • Meter Information: View the meter model, type, serial number and current software version. •...
  • Page 47  when the factory reset is completed. Note: If Secure Mode is enabled, not all options may be accessible to users in the Information & Help list, depending on each user’s meter permissions. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 48: Chapter 4

    Chapter 4 Calibration 4.1 pH Calibration Refer to the 3.9 Information & Help section and Meter Tours to access onscreen tours that explain how to use important meter functions including pH calibration. Refer to the 3.2.2 pH Settings section and Alarm Settings to use the Calibration Reminder Alarm to set a reminder if calibration is not performed within the selected time interval.
  • Page 49: Ph Calibration Procedure

    (4, 3, 2 or 1), only the Next button will appear for each calibration point until the set number of points is completed, at which point the Done button will appear. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 50: Ph Value Vs. Temperature

    4.1.3 pH Value vs. Temperature The following tables list theoretical values for pH buffers at various temperatures. Actual values shown on the meter may vary due to rounding deviations from the actual pH and temperature measurements. USA Buffer Set (Orion pH Buffers) Temp.
  • Page 51 4.120 6.950 9.120 4.135 6.945 9.080 4.150 6.940 9.040 4.165 6.940 9.000 4.180 6.940 8.960 4.200 6.940 8.920 4.220 6.940 8.880 4.245 6.940 8.840 4.270 6.940 8.800 4.295 6.940 8.760 4.320 6.940 8.720 Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 52: Orp Calibration

    4.2 ORP Calibration Refer to the 3.2.2 pH Settings section and Alarm Settings to use the Calibration Reminder Alarm to set a reminder if a calibration is not performed within the selected time interval. 4.2.1 ORP Calibration Overview The ORP calibration is a millivolt (mV) offset adjustment. Oxidation reduction potential (ORP) or redox is useful as a relative indicator of the oxidizing or reducing nature of a sample solution.
  • Page 53 This may be helpful if the electrode is responding more slowly than usual. Tap force stabilize text to accelerate the reading stability (may give less accurate calibration results). Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 54: Temperature Calibration

    4.3 Temperature Calibration 4.3.1 Temperature Calibration Overview The thermistor sensor used for automatic temperature compensation (ATC) and measurements is both accurate and stable, so frequent calibration is not required. If desired, perform a temperature calibration to calibrate the ATC sensor when it is connected to the meter. A 1 to 2 point temperature offset calibration can be performed, up to ±5.0°C for each calibration point.
  • Page 55: Chapter 5

    Meter Tours to access onscreen tours that explain how to use important meter functions including measurements. The Orion Pro Star PH211 meter is capable of measuring pH, mV or ORP with temperature Refer to the 2.5 Touchscreen Functions section for details on the display components and icons.
  • Page 56: Measure Mode Read Types

    View Active Calibration Details Mode Example Calibration Details Notes The pH electrode status icon represents the average slope of the most recent calibration. The icon is green when the slope is 92.0% to 102.0%. The icon is orange when the slope is Calibration 82.0% to 91.9% or 102.1% to 114.0%.
  • Page 57 • Tap the Stop button or leave the measure mode to stop taking timed measurements. • The Orion stirrer probe will turn on automatically when the Start button is tapped. When on, tap Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 58: Sample Id

    the Stirrer button to manually turn off the Orion stirrer probe. The Orion stirrer probe will turn off automatically when the Stop button is tapped. Timed, Waiting Timed, Reading Stabilizing Timed, Reading Stable 5.1.2 Sample ID When using Sample ID to identify samples with measurements, refer to the 3.2.2 pH Settings section and Sample ID setting to turn on sample ID, set the type of sample ID and enter the sample ID.
  • Page 59: Stirrer Button

    Stop button is tapped. Press and hold the Stirrer button to access and change the stirrer control settings. Note: The Stirrer button is hidden if the stirrer function is turned off. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 60: General Measurement Procedure

    5.2 General Measurement Procedure Refer to the electrode manuals for detailed instructions on electrode preparation, calibration and usage. Make sure the electrodes are fully prepared, properly calibrated and working correctly before taking sample measurements. Refer to the 3.2.1 Instrument Settings section and Read Type settings to define how measurements are performed and when measurements are saved and exported.
  • Page 61: Data Viewing

    • When in Secure Mode, the meter will save up to 20 calibration logs and prompt the Administrator and Level 1 accounts to export and delete calibration logs when the limit is reached. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 62: Chapter 6

    Chapter 6 Data Exporting and Importing 6.1 Data Transfer Export live measurements, completed calibrations, data logs, calibration logs, methods, user lists (Secure Mode only) and audit trails (Secure Mode only) from the meter to a USB flash drive, computer or Orion ink printer (catalog number STARA-106). To successfully export data: Set all required data transfer settings in the Instrument Settings menu.
  • Page 63: Connect The External Device

    Printer Setting Baud Rate : 115200 bps Baud Rate : 9600 bps Data bits: 8 Data bits: 8 Parity: None Parity: None Stop bits: 1 Stop bits: 1 Flow Control: None Flow Control: None Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 64: Verify The Data Destination

    6.1.3 Verify the Data Destination USB Flash Drive For data transfer using a USB flash drive, perform a test export of data from the meter to the USB flash drive. Remove the USB flash drive from the meter, insert it in a computer and open the test export on the computer.
  • Page 65 TIMEMODE <CR> TIME MODE = 1 (PM), 2 (AM) or 3 (24 Hour) SETMODE MODE Sets the meter measurement mode. MODE = PH or SETMODE PH <CR> GETMODE <CR> Prints the active measurement mode Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 66: Meter Software Update Procedure

    6.3 Meter Software Update Procedure Note: Back up all data before performing the meter software update procedure, as it will be erased when the meter software update is completed. Power on the meter. Download the latest Orion Pro Star meter software at www.thermofisher.com/orionsoftware. Unzip/extract the Orion Pro Star meter software file.
  • Page 67: Chapter 7

    Perform a factor reset on the meter. o Replace the electrode. • pH Electrode Icon: o Troubleshoot the pH buffers and pH electrode. o Replace the pH buffers. o Clean and condition the pH electrode. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 68: Measurement Troubleshooting

    o Replace the pH electrode as needed. 7.2 Measurement Troubleshooting Reading is unstable, slow to stabilize: • Consider changing the displayed measurement resolution. For example, some pH measurements only require 0.01 pH reporting. Changing the pH resolution from 0.001 to 0.01 will reduce the amount of time needed for the reading to stabilize.
  • Page 69: Technical Support

    ORP standard into a clean, dry beaker and repeat the calibration. 7.4 Technical Support Contact our Orion Technical Support team at 1-800-225-1480, +1-978-232-6000 or WLP.TechSupport@thermofisher.com. You will need the meter model, meter serial number and meter software revision information. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 70: Chapter 8

    Chapter 8 Meter Info 8.1 Meter Specifications Range -2.000 to 20.000 pH Resolution 0.1, 0.01 or 0.001 pH Relative accuracy ± 0.002 pH Calibration points 1 to 5 points Calibration Automatic buffer recognition, manual entry option method USA (1.68, 4.01, 7.00, 10.01, 12.46) DIN (1.68, 4.01, 6.86, 9.18, 12.46) Calibration buffer Pure Water Buffers (4.10, 6.97, 9.15)
  • Page 71 USB flash drive, computer or Orion ink printer Data output Report, CSV or print format Data log 10,000 data sets with time and date Data log export Customizable options for basic (short), intermediate, advanced or all data Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 72 Calibration log 20 per parameter with time and date Calibration log Customizable options for basic (short), intermediate, advanced or all data export Calibration 1 alarm per parameter reminder Calibration alarm 1 hour to 30 days or off interval Audible signal User set for button tap, notifications and alarms, stable reading Maintenance Up to 5 electrode maintenance reminders and up to 5 solution maintenance...
  • Page 73: Meter Dimensions

    Note: Specifications subject to change without notice. All relative accuracy values are influenced by the displayed least significant digit and should include ± 1 LSD for all relative accuracy calculations. 8.2 Meter Dimensions Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 74: Warranty

    8.3 Warranty This meter is supplied with a warranty against significant deviations in material and workmanship for a period of 2 years from date of purchase. Electrode warranties are separate from the meter and differ based on the selected electrode. If meter repair is necessary within the designated warranty period and has not been the result of abuse or misuse, please contact the Technical Support Team at WLP.TechSupport@thermofisher.com...
  • Page 75: Chapter 9

    It is important to use only accessories as supplied or recommended by Thermo Fisher Scientific. 9.3 Electromagnetic Compatibility The meter requires special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information provided in this instruction manual. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 76: Fcc Statement (Usa)

    Recommendations for actions to ensure the product remains safe regarding electromagnetic disturbances: Do not modify the product beyond the intended user-adjustable settings. It is important to use only accessories as supplied or recommended by Thermo Fisher Scientific. The equipment should be visually inspected regularly for damaged cables and connectors. Damaged cables should be replaced.
  • Page 77: Environmental Compliance

    EN 63000:2018. This confirms that any potential trace contamination levels of the substances listed above are below the maximum level set by the latest regulations or are exempt due to their application. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 78: Weee Compliance

    Ulteriori informazioni sulla conformità di Thermo Fisher Scientific con queste Direttive, l’elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodotti Thermo Scientific che possono essere utili alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito http://www.thermofisher.com.
  • Page 79: Proper Product Disposal

    Safety Data Sheets (SDS), or regulations applicable to your country and region. • This product has materials and components that may be recycled or reused according to local guidelines and regulations. • Clean up any chemical safety hazards using appropriate methods. Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 80: Chapter 10

    (8302BNUMD); pH 4, 7, 10 buffers and rinse solution pouches; pH electrode storage solution and cleaning solution; pH electrode storage sleeve and base Orion Pro Star PH211 pH/ORP bench meter durable kit; includes meter with stand PSTAR2116 (PSTAR2110); ROSS Ultra Triode refillable epoxy-body pH/ATC electrode (8157BNUMD);...
  • Page 81: Electrode And Solution Ordering Info

    Standard All-in-One pH buffer kit: 475 mL each pH 4.01, 7.00 and 10.01 buffers 910199 and pH electrode storage solution ROSS All-in-One pH buffer kit: 475 mL each pH 4.01, 7.00 and 10.01 buffers 810199 and ROSS pH electrode storage; 60 mL electrode cleaning solution Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 82 Description Catalog Number pH 1.68 buffer, 475 mL 910168 pH 4.01 buffer, 475 mL 910104 pH 6.86 buffer, 475 mL 910686 pH 7.00 buffer, 475 mL 910107 pH 9.18 buffer, 475 mL 910918 pH 10.01 buffer, 475 mL 910110 pH 12.46 buffer, 475 mL 910112 pH 4.01 buffer, 25 individual-use pouches 910425...
  • Page 83: User Notes

    User Notes ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ Orion Pro Star PH211 pH/ORP Bench Meter...
  • Page 84 Orion Pro Star PH211 pH/ORP Bench Meter Scientists have trusted Thermo Scientific Orion Lab Products for over 60 years to provide them with accurate and reliable analytical testing instruments in the lab and in the field. Since 1962, Thermo Scientific Orion products have set the standard for quality in electrochemistry. We offer instruments for analytical and optical measurement that allow scientists to solve some of the most challenging issues facing society today.

Table of Contents