8. Bluetooth
Function
This device can be used in conjunction with a smartphone running
the «Microlife Connected Health+» App.
®
Bluetooth
can be activated manually by pressing and holding the
START/STOP button AN for about 4 seconds until the Bluetooth
symbol begins flashing BS.
®
Bluetooth
is automatically activated BS after a measurement has
been completed.
Bluetooth pairing and set up is required to use this device
with the «Microlife Connected Health+» App. It is recom-
mended to manually activate Bluetooth
«Microlife Connected Health+» App to synchronize the date
and time settings between the device and the smartphone.
In case of difficulties connecting the device with the «Micro-
life Connected Health+» App, try to reset the Bluetooth
connection of this device to default settings:
Press and hold the START/STOP button AN for 8 seconds
until «CL Pr» is displayed. This clears the Bluetooth
pairing records on the device and reset it to factory settings.
®
Bluetooth
automatically activates after the reset.
For more detailed information visit www.microlife.com/connect.

9. PC-Link Functions

This device can be used in conjunction with a personal computer
(PC) running the Microlife Blood Pressure Analyzer+ (BPA+) soft-
ware. The memory data can be transferred to the PC by
connecting the monitor via a cable.
If no download-voucher and cable is included download the BPA+
software from www.microlife.com/software and use a Micro-USB
cable.

10. Error Messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «Err 3», is displayed.
Error
Description Potential cause and remedy
«Err 1»
Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat the
measurement.*
6
Error
«Err 2»
BK
«Err 3»
BT
«Err 5»
®
and connect to the
«Err 6»
®
®
«HI»
«LO»
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.

11. Safety, Care, Accuracy Test and Disposal

 Follow instructions for use. This document provides important
Description Potential cause and remedy
Error signal During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
Abnormal
An adequate pressure cannot be gener-
cuff
ated in the cuff. A leak may have occurred.
pressure
Check that the cuff is correctly connected
and is not too loose. Replace the batteries
if necessary. Repeat the measurement.
Abnormal
The measuring signals are inaccurate and
result
no result can therefore be displayed. Read
through the checklist for performing reli-
able measurements and then repeat the
measurement.*
AFIB/MAM
There were too many errors during the
Mode
measurement in AFIB/MAM mode,
making it impossible to obtain a final
result. Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
Pulse or cuff
The pressure in the cuff is too high (over
pressure too
299 mmHg) OR the pulse is too high (over
high
200 beats per minute). Relax for 5 minutes
and repeat the measurement.*
Pulse too
The pulse is too low (less than 40 beats
low
per minute). Repeat the measurement.*
Problem
If any problem occurs with the Bluetooth
with Blue-
connection, the Bluetooth
®
tooth
blinks rapidly for approximately 10
connection
seconds. To solve the problem, please
visit www.microlife.com/connect.

Safety and protection

product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
®