Functional Principle, Technical Data
2.3
Technical Data
Controller modell
Resolution static
Resolution dynamic
Frequency response
Linearity
Sensitivity
Long-term stability
Temperature stability
Sensor
Operating
Sensor cable
temperature
Controller
Power supply
analog
Output
digital
Storage temperature
Sensor
Protection
class
Controller
Shock
Vibration
Weight
FSO = Full Scale Output
capaNCDT 61xx/IP
DT6110/IP/U
50 ppm FSO/K (+10 ... +40 °C);
type CSF -40 ... +100 °C | type CSE: -50 ... +200 °C
permanent: -20 ... +80 °C, up to 10,000 h: -20 ... +100 °C
9 ... 28 VDC
(max. 1.4 W)
0 ... 10 V
(short circuit proof)
max. 500 Ohm load
optional: ±5 V, 10 ... 0 V
---
20 g / half sine 5 ms in XYZ axis / 1000 shocks per axis (DIN-EN 60068-2-29)
10 g / 10 ... 2000 Hz in XYZ axis / 10 cycles per axis (DIN-EN 60068-2-6)
DT6110/IP/I
DT6120/IP/U
0.01 % FSO (2 Hz)
0.02 % FSO (1000 Hz)
1 kHz (-3 dB)
≤ ±0.1 % FSO
≤ ±0.1 % FSO
0.02 % FSO/month
100 ppm FSO/K (-20 ... +10 °C and +40 ... +60 °C)
-20 ... +60° C
11 ... 28 VDC
9 ... 28 VDC
(max. 2 W)
(max. 1.6 W)
4 ... 20 mA
(short circuit proof)
optional: ±5 V, 10 ... 0 V
---
24 bit measuring values, max. 2 kSamples (adjustable)
-20 ... +75 °C
IP 40, IP 54 (functionality reduced)
IP 68
165 g
DT6120/IP/I
11 ... 28 VDC
(max. 2.2 W)
0 ... 10 V
max. 500 Ohm load
RS485, 230400 Baud (adjustable),
4 ... 20 mA
Seite 14
Need help?
Do you have a question about the capaNCDT 6110 IP and is the answer not in the manual?