Functional Principle, Technical Data
2.2.1
For this measuring system, several sensors can be used.
The capacitive measuring process is area-related. A minimum area, see
Fig.
Sensor model
CSE05
CSE1
CSE1,25
CSE2
CSE3
Sensor model
CSF2
CSF4
CSF6
Fig. 5 Sensors for electrical conducting targets (metals)
capaNCDT 61xx/IP

Sensors

In order to obtain accurate measuring results, keep the surface of
the sensor clean and free from damage.
5, is required depending on the sensor model and measuring range.
CSE05/M8
CSE1,25/M12
CSE2/M16
CSE3/M24
CSF2-CRg4,0
CSF4-CRg4,0
CSF6-CRg4,0
Measurement direction
Measuring range,
nominal
0.5 mm
1 mm
1.25 mm
2 mm
3 mm
Measuring range,
nominal
2 mm
appr. 50.5 x 14.0 mm ≥ 0.75 mm
4 mm
appr. 90.5 x 17.5 mm ≥ 0.75 mm
6 mm
appr. 127.3 x 25 mm ≥ 0.75 mm
Measuring
surface
sensor
Min. target
diameter
Min. target
diameter
6 mm
8 mm
10 mm
14 mm
20 mm
Min. target
Min gap
size
width