Work Area Air Cleanliness Test
(optional)
If the customer wants to verify air cleanliness in the work area, use the following
procedure:
1. Mark test points at the geometric center of the work surface, and midway
between the center test point and each side wall.
2. Place a particle counter pickup a test point, at a height of 16 inches above the
work surface, pointing into the airflow, as shown in Figure 5-23. Take three
samples of 1.0 ft
3. Average the results.
Acceptance
The average count should be less than 100 particles 0.5um in size per ft
. Repeat for each test point.
3
Product Service 1-800-522-7658
Figure 5-23
3
.