Calibration Procedure-
NOTE: YOU MUST FOLLOW THESE INSTRUCTIONS EXACTLY, OR THE SENSOR
Note: You will need to access the certifiers menu to calibrate the airflow
sensor.
Note: To properly calibrate the air flow sensor, a qualified certifier must
verify the inflow and downflow values.
With the cabinet operating:
Menu
8. Press the
9. Press the
button until the
▼
(it will turn yellow).
OK/Mute
10. Press
11. Press the
button until the
▼
highlighted (it will turn white).
OK/Mute
12. Press
13. You will see a
about to alter the BSC's settings.
OK/Mute
14. Press
NOTE: IF THE UNIT IS CONNECTED TO AN EXHAUST SYSTEM, YOU MUST
STOP ALL AIR FLOW OVER THE SENSOR DURING THE INITIAL ZEROING
OF THE SENSOR - DO THIS BY SHUTTING OFF THE EXHAUST SYSTEM, OR
BLOCKING THE SENSOR INTAKE WITH TAPE.
WILL NOT OPERATE PROPERLY!
button.
Services
to enter the Services menu screen.
Calibration
to start the calibration procedure.
warning screen
alerting you that you are
to input the password.
Product Service 1-800-522-7658
option is highlighted
option is
MAIN MENU
TM
1. MyLogic
2. Settings
3. Service
or
to choose,
OK to select, MENU to escape
Service
Current Settings
Restore Default
Diagnostics
Calibration
Configuration
or
to choose,
OK to select, MENU to return
Calibration
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
OK to proceed, MENU to return
145