9. The first screen allows you to set the Type of cabinet. The only difference
between the Type A1 and Type A2 configuration setting is that the airflow
velocity limits for normal operation of the optional airflow
sensor (because the Type A1 has a nominal inflow of 85
FPM, and the Type A2 is 105 FPM). If the air velocity goes
out of range, the alarm will be displayed, but the cabinet will
continue to run. Selecting Type B2 will cause an inflow
alarm if the inflow velocity drops below the alarm setpoint,
and will result in the cabinet blower stopping. Use the
buttons to select the cabinet Type, and then press
▲or▼
OK/Mute.
10. The second screen allows you to configure the cabinet for a
canopy or not. Select the "a canopy" option.
11. This screen configures the unit for a UV light or not. After
your selection is chosen, and you press OK/Mute, the unit
will shut off and then reset in the new configuration.
Product Service 1-800-522-7658
Configuration
Type A Canopy Option
This cabinet has:
a canopy
no canopy
or
to choose,
OK to select, MENU to escape
UV Settings
This Logic is configured:
to have a UV light
to NOT have a UV light
or
to choose,
OK to select, MENU to return
141
Need help?
Do you have a question about the Purifier Logic+ 30132 and is the answer not in the manual?