Configuring the Logic+
In the first MyLogic screen, use the
buttons on the touchpad to highlight
Logic for use
-it will turn white when selected. Press
OK/MUTE
to enter the first configuration screen:
The first screen gives you the option of activating
the Smart-Start option for the blower; if you want
the cabinet blower to start every time you raise the
start
sash, select
, and then press
start
is selected, then the blower must be manually
started from the keypad. When
pressed, the next configuration screen will appear.
The next screen gives you the option of activating
the Smart-Start option for the fluorescent light; if
you want the lights to turn on every time you raise
turn on
the sash, select
stay off
If
is selected, then the lights must be
manually turned on from the keypad. When
OK/MUTE
is pressed, the next configuration screen
will appear.
If you want the cabinet blower to run slowly,
maintaining reduced airflows every time you close
go into Night-Smart mode
the sash, select
OK/Mute
stop
press
. If
blower will stop when the sash is closed. When
OK/MUTE
is pressed, the next configuration screen
will appear.
If your Logic+ is configured for a UV light, then you
will see the next two screens; if you want the UV
lamp to turn on every time you close the sash,
go into Night-Smart mode
select
OK/Mute
stay off
. If
will not turn on when the sash is closed. When
OK/MUTE
is pressed, the final configuration
screen will appear.
and
▲
configure my
OK/Mute
OK/MUTE
OK/Mute
, and then press
and then
is selected, then the
, and then press
is selected, then the UV light
Product Service 1-800-522-7658
▼
not
. If
is
.
65