Labconco Purifier Logic+ 30132 Technical Manual And Specifications page 170

Table of Contents

Advertisement

PROBLEM
Fluorescent light
is dim or
flickering
UV light not
working
UV light is dim or
flickering
Airflow Alert
goes off and/or
there is a slight
decrease in filter
life remaining
gauge
170
CAUSE
Lamp(s) are
defective
Lamp wiring is
disconnected
Defective lamp
ballast
Sash is open
Lamp is defective
Lamp wiring is
disconnected.
Defective lamp
ballast
Keypad
disconnected or
defective
Lamp is defective
or is at end of
operating lifetime.
Lamp wiring is
disconnected
Defective lamp
ballast
HEPA filter loading
Blockage of the
return air slots or
grille
Blockage of the
Product Service 1-800-522-7658
CORRECTIVE ACTION
Replace defective lamp(s)
Inspect lamp wiring.
Replace lamp ballast.
Close sash – UV light will not work
with the sash open.
Replace defective lamp.
Inspect lamp wiring.
Replace lamp ballast.
Run keypad diagnostics and check
connections.
Replace defective or worn out lamp.
Inspect lamp wiring.
Replace lamp ballast.
The gauge reading steadily
decreases as the Cabinet is used.
Check all return air slots and grilles
to ensure that they are not blocked
or restricted.
Ensure that the exhaust outlet is