Quickstart
How do I run a diagnostic on the sash position switches?
Menu
1. Press the
2. Press the
▼
button until the
turn yellow).
OK/Mute
3. Press
▼
4. Press the
button until the Diagnostic
highlighted (it will turn white).
OK/Mute
5. Press
6. Press the
▼
button until the Diagnostic
highlighted (it will turn white). Press
7. Lower the sash, and note the status line – it should properly
report the sash as open too high, or open, or closed when the
sash is in the corresponding position.* Press
the diagnostic.
*Early production models of Type A Logic+ units may display a sash height alarm screen
when the sash is raised above its operating position. This is normal.
button.
Services
option is highlighted (it will
to enter the Services menu screen.
to enter the Diagnostic submenu.
OK/Mute
Product Service 1-800-522-7658
Diagnostic
option is
Sash sensors
option is
.
OK/Mute
to exit
MAIN MENU
TM
1. MyLogic
2. Settings
3. Service
or
to choose,
OK to select, MENU to escape
Service
Current Settings
Restore Default
Diagnostics
Calibration
Configuration
or
to choose,
OK to select, MENU to return
Diagnostics
Sash sensors
Keypad -Relays
Canopy Alarm
Airflow Sensor
or
to choose,
OK to select, MENU to escape
Diagnostics
Lower the sash from full open
to closed
Note the sash status below:
Sash Status:
The sash is open too high
The sash is open
The sash is closed
OK to return
15