5. Using the screws, flat washers and locking nuts provided, secure the UV ballast to
the top of the electronics module.
6. Wire the UV ballast into the wiring harness, following the wiring diagrams located in
Figure 2a-c.
7. Reattach the electronics module to the top of the cabinet before reconnecting the
electrical power.
8. Raise the sash, and install the UV lamp by rotating the lamp pins into the sockets.
9. For all Logic Models:
Lower the sash to fully closed position, plug in the unit, and turn the system
power switch to the "UV" position. The UV lamp will illuminate. Refer to the
user's manual for further information regarding the proper operation and
maintenance of the UV light.
For All Logic+ models:
Lower the sash to fully closed position, plug in the unit, and do the following:
Menu
12. Press the
13. Press the
▼
button until the
turn yellow).
OK/Mute
14. Press
15. Press the
▼
button until the
(it will turn white).
OK/Mute
16. Press
warning screen
17. You will see a
alter the BSC's settings.
OK/Mute
18. Press
19. When requested for the password press
Timer
Timer
,
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Configuration
to alter the configuration.
alerting you that you are about to
to input the password.
OK/Mute
then
.
Product Service 1-800-522-7658
option is highlighted
Light
UV light
,
,
MAIN MENU
TM
1. MyLogic
2. Settings
3. Service
or
to choose,
OK to select, MENU to escape
Service
Current Settings
Restore Default
Diagnostics
Calibration
Configuration
or
to choose,
OK to select, MENU to return
Configuration
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
OK to proceed, MENU to return
Configuration
PASSWORD:
151