UV Settings

For models equipped with the optional UV light the Logic+ has an integral UV light
maintenance system. It allows you to define how many hours you want the UV lamp to
operate before receiving a reminder to replace it, a way to monitor how many hours the
lamp has been on, and the means to reset the UV lamp hourmeter.

UV Lamp Hourmeter

This display only shows how many hours the UV lamp has been lit, and how many hours
remain until you will receive a warning to replace the lamp.
From the Settings menu screen, use the
buttons on the touchpad to highlight the
option-it will turn white when selected. Press
enter the UV Settings screen. Using the
on the touchpad, highlight UV Hourmeter-it will turn white
when selected, and then press
then return to the UV Hourmeter screen.
This screen displays how many hours the UV lamp has
operated, and how many hours of operation remain before
replacement is recommended. When finished with this
MENU
screen, press
screen.

Reset UV Lamp Hourmeter

This option lets you reset the UV hourmeter to 0 hours
whenever the lamp has been replaced.
This screen allows you to reset the UV Hourmeter to 0
OK/Mute
hours if you press
you to the first UV Settings screen without resetting the
hourmeter.

Change UV Lamp Life

In this screen, you can set the number of operating hours before receiving the replace
UV lamp warning. For most UV lamps, the output of UV light decreases at a constant
rate. Typically, after 6,000 hours of operation the lamp will output 80%. This option
allows you to set operational life of the UV lamp, in 100 hour increments.
From the UV Settings menu screen, use the
buttons on the touchpad to highlight the
it will turn white when selected. Press
the Set UV life screen. Use the
or decrease the lamp life in 100 hour increments, until the
desired value is displayed, and then press
screen will then return to the UV Settings screen.
68
UV Settings
OK/Mute
. The screen will
to return to the first UV Settings
MENU
. Pressing
Set UV Life
OK/MUTE
and
buttons to increase
Product Service 1-800-522-7658
and
OK/MUTE
to
and
buttons
will return
and
option-
to enter
OK/Mute
. The