Canopy Alarm - Labconco Purifier Logic+ 30132 Technical Manual And Specifications

Table of Contents

Advertisement

Quickstart
How do I run a diagnostic on the Canopy Alarm?
Menu
1. Press the
2. Press the
button until the
turn yellow).
OK/Mute
3. Press
4. Press the
button until the
will turn white).
OK/Mute
5. Press
6. Press the
button until the Diagnostic
highlighted (it will turn white). Press
7. The display shows the status of the Canopy Inlet Valve. During
normal operation, it will be open, and during a canopy alarm, it
will be closed. NOTE: The canopy door function is OPPOSITE OF
THE MAGNETIC SWITCH FUNCTION! – WHEN THE INLET VALVE
IS OPEN, THE MAGNETIC SWITCH IS CLOSED.
Quickstart
How do I run a diagnostic on the Airflow Sensor?
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Diagnostic
to enter the Diagnostic submenu.
OK/Mute
Product Service 1-800-522-7658
option is highlighted (it

Canopy Alarm

option is
.
MAIN MENU
TM
1. MyLogic
2. Settings
3. Service
 
or
to choose,
OK to select, MENU to escape
Service
Current Settings
Restore Default
Diagnostics
Calibration
Configuration
 
or
to choose,
OK to select, MENU to return
Diagnostics
Sash sensors
Keypad - Relays
Canopy Alarm
Airflow Sensor
 
or
to choose,
OK to select, MENU to escape
17