Quickstart
How do I recalibrate the
blower speed
filter life gauge
airflow sensor reading
11. If this is an initial certification or a recertification after HEPA filter
replacement, select
annual recertification, select
want to set the filter gauge at a preset value, select
to new setting
. Then press
12. The Type A units with airflow sensor will ask you to now
calibrate the average downflow to match the value you obtained
when you measured it. After the downflow is calibrated, and you
OK/Mute
press
with the new settings.
13. If the unit is a Type A without an airflow sensor, after the filter
gauge option is selected and you press
will power down, and then restart with the new settings. If the
Unit is a Type B2 or a Type A with an airflow sensor, you will
then be asked to calibrate the inflow sensor to match the value
you obtained when you measured the inflow. After the inflow is
calibrated, and you press
down, and then restart with the new settings.
14. For the Type B2, the unit will now ask you to reduce the
exhaust volume to 80% of the nominal value. When you have
done this, press
point for the exhaust flow. If you press OK/Mute before reducing
the flow to 80%, the Logic+ will assume the alarm setpoint is
equal to the nominal airflow, and the unit will return an Alarm-
Inflow Screen.
with the new settings. Be sure to reset the exhaust damper to its
nominal setting.
of the Logic+? - Continued
Reset Filter Gauge to 100%
Leave Gauge unchanged
OK/Mute
.
, the Logic+ will power down, and then restart
OK/Mute
, the Logic+ will power
OK/Mute
.
This step is critical-it sets the alarm
The Logic+ will power down, and then restart
Product Service 1-800-522-7658
. If this is an
. If you
Set Gauge
OK/Mute
, the Logic+
total
Calibration
Filter Gauge Adjustment
Reset Filter Gauge to 100%
Leave Gauge unchanged
Set Gauge to new setting
or
to change filter gauge,
OK to select, MENU to finish
Calibration
Inflow Display
The Inflow displayed
value is: 102 FPM
or
to change display value,
UV Light to reset to 105 FPM,
OK to select, MENU to finish
Calibration
Inflow Alarm Setpoint
Reduce the exhaust flow to
80% of nominal volume. Press
OK when you are ready to fix
the setpoint.
OK to select, MENU to finish