Labconco Purifier Logic+ 30132 Technical Manual And Specifications page 147

Table of Contents

Advertisement

Confirming Calibration Procedure-
With the cabinet operating:
Menu
1. Press the
2. Press the
button until the
(it will turn yellow).
OK/Mute
3. Press
4. Press the
button until the
highlighted (it will turn white).
OK/Mute
5. Press
6. Press the
button until the
highlighted (it will turn white).
OK/Mute
7. Press
screen.
8. The display shows the following values in millivolts -
a. "Zero" – should be approximately 90-110.
b. Alarm (approximately 85% of Full scale – calculated
by the cabinet)
c. Full scale - the value for inflow set during
calibration.
NOTE: There should be approximately 100 millivolts or higher difference
between the zero and operational (full scale) points. Differences
significantly lower than 100 mV may result in erratic air velocity values
being displayed
button.
Services
to enter the Services menu screen.
Diagnostics
to enter the Diagnostics submenu screen.
Airflow Sensor
to enter the Airflow Sensor submenu
.
Product Service 1-800-522-7658
option is highlighted
option is
option is
MAIN MENU
TM
1. MyLogic
2. Settings
3. Service
 
or
to choose,
OK to select, MENU to escape
147