Quickstart
How do I recalibrate the
blower speed
filter life gauge
airflow sensor reading
Menu
1. Press the
2. Press the
▼
button until the
turn yellow).
OK/Mute
3. Press
4. Press the
▼
button until the
will turn white).
OK/Mute
5. Press
warning screen
6. You will see a
alter the BSC's settings.
7. Press
OK/Mute
8. When requested for the password press
Timer
Timer
,
9. If the password is properly entered, the first certification screen
will display the blower motor's PWM setting.
10. Press the
▲
or
speed as needed. When the speed is set properly, press the
OK/Mute
button to lock the new blower speed setting, and
continue to the filter life screen.
10
of the Logic+?
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Calibration
to start the calibration procedure.
alerting you that you are about to
to input the password.
OK/Mute
then
.
▼
button to increase or decrease the blower
Product Service 1-800-522-7658
option is highlighted (it
Light
UV light
,
,
MAIN MENU
TM
1. MyLogic
2. Settings
3. Service
or
to choose,
OK to select, MENU to escape
Service
Current Settings
Restore Default
Diagnostics
Calibration
Configuration
or
to choose,
OK to select, MENU to return
Calibration
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
OK to proceed, MENU to return
Calibration
PASSWORD:
Calibration
Motor Speed Adjustment
The ECM setting is: 80
The current RPM is: 1050
The average RPM is: 1020
or
to change motor speed,
OK to select, MENU to finish