Configuring the Logic+

In the first MyLogic screen, use the
buttons on the touchpad to highlight
Logic for use
-it will turn white when selected. Press
OK
to enter the first configuration screen:
The first screen gives you the option of activating
the Smart-Start option for the blower; if you want
the cabinet blower to start every time you raise the
sash, select start, and then press OK. If
selected, then the blower must be manually started
from the keypad. When
configuration screen will appear.
The next screen gives you the option of activating
the Smart-Start option for the fluorescent light; if
you want the lights to turn on every time you raise
the sash, select
turn
on, and then press OK. If
off
is selected, then the lights must be manually
turned on from the keypad. When
next configuration screen will appear.
If you want the cabinet blower to run slowly,
maintaining reduced airflows every time you close
the sash, select
go into Night-Smart mode
press OK. If
stop
is selected, then the blower will
stop when the sash is closed. When
the next configuration screen will appear.
If your Logic+ is configured for a UV light, then
you will see the next two screens; if you want the
UV lamp to turn on every time you close the sash,
select
go into Night-Smart
If
stay off
is selected, then the UV light will not turn
on when the sash is closed. When
final configuration screen will appear.
30
and
configure my
not start
OK
is pressed, the next
stay
OK
is pressed, the
and then
OK
is pressed,
mode, and then press OK.
OK
is pressed, the