2. In the Windows "Start" menu, enter "Services" and check the state of the following
services.
Figure 8-9: R&S TSMA6 Services
If both services are running, continue with
3. Start the services, that are not running.
a) Select the service to be started.
b) Open the context menu with a right-mouse click.
c) Start the service.
If the services could not be started, reinstall the firmware (see
ware
4. Try again to start the web GUI locally on the R&S TSMA6 via Internet Explorer.
User Manual 4900.8057.02 ─ 06
R&S TSMA6 running?
R&S WebServer running?
Update",
on page 53).
Troubleshooting
Web GUI Locally Not Accessible
step
4.
Chapter 7.1, "Firm-
113