Monitoring System; Overview; Power-On Screen - Dräger Narkomed 2B Operator's Instruction Manual

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

Overview

Power-On Screen

In addition to monitoring clinical parameters, the Narkomed 2B
performs diagnostic self-tests every time the machine is turned on. After
the initial power-on screen is displayed, the monitor screens are
displayed. This section of the manual describes the power-on screen and
explains how to work with the monitoring screens and their controls.
When you turn the SYSTEM POWER switch to the ON position, the
Narkomed 2B performs extensive self-tests on its internal hardware. As
these diagnostics are performed, each test and its result (PASS or FAIL)
appear on the left-hand display screen, indicating the status of various
components of the monitoring system.
VIDEO TEST
FIRMWARE TEST
MEMORY TEST
TIMERS TEST
ANALOG TEST
AUDIO TEST - PRIMARY
BACK UP
SERIAL I/O TEST
CLOCK TEST
BACKUP MEMORY TEST
AC POWER TEST
RESERVE POWER TEST
FUNCTIONAL
OP20056
At the end of the self-diagnostics, one of three possible conclusions to the
self-tests is posted on the left-hand screen:
FUNCTIONAL
CONDITIONALLY
FUNCTIONAL
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
Every component of the monitoring system is in
satisfactory operational order. Normal operation of
the machine continues.
A noncritical fault was detected, such as a speaker
failure. You can use the Narkomed 2B, but first
note the test that failed. Report the problem to a
North American Dräger qualified technical service
representative when convenient.
Section 5 - Operation

Monitoring System

NARKOMED 2B
VERSION x.xx - DIAGNOSTICS
COPYRIGHT, NAD INC. 1987-94
5-9-1