Vaporizers - Dräger Primus SW 4.5n Instructions For Use Manual

Anesthesia workstation
Table of Contents

Advertisement

Getting started

Vaporizers

NOTE
Before operating the vaporizer, pay special atten-
tion to the instructions for use of the vaporizer be-
ing used. Note especially the vaporizer flow limits.
The vaporizer being used must comply with stand-
ard ISO 8835-4 or ISO 80601-2-13. If the internal
patient-gas measurement module fails, an inde-
pendent gas measurement system complying with
ISO 21647 or ISO 80601-2-55 must be used.
The Vapor 2000 is shown and described here.
A
D
0
C
B
76
A
0
C
B
For the Dräger Interlock 2 System:
Vaporizers are mounted in a level position and
seated securely on the mounts.
1 Locking levers (A) point to the left = locked po-
sition.
2 Check the sight glass (B) and ensure an ade-
quate filling level.
3 Handwheel set to 0 and button (C) engaged.
4 Check the interlock mechanism.
Move the selector lever (D) to the left to lock the
left vaporizer. Turn the handwheel on the right
vaporizer to a position other than 0, and make
sure the left vaporizer remains locked in its 0
position. Repeat test for other vaporizer.
5 Turn both handwheels to 0 positions.
Instructions for use Primus SW 4.5n