Parameterizing The Device; Submenu "Set Corr": Offset/Zero Point Correction And Slope Correction Of The Temperature Measurement - JUMO 202710/30 Operating Manual

Handheld meter for conductivity (cr), specific electrical resistivity, tds, and salinity
Hide thumbs Also See for 202710/30:
Table of Contents

Advertisement

6.4

Parameterizing the device

6.4.1 Submenu "SEt Corr": offset/zero point correction and slope cor-
rection of the temperature measurement
Overview of the parameters
Parameter "OFFS": offset/zero point correction of the temperature measurement
After calling up the setting menu with
"SEt Corr", use
to select the parameter to be configured.
SAL
mg/l kOhm
mS
µS
Lin
nLF
AL Logg
%/K 1/cm
Parameters
(1)
"OFFS": offset/zero point correction of the temperature measurement, see
page 29
(2)
"SCAL": slope correction of the temperature measurement, see page 30
Use the
and
Menu/parameter
Press the
key to save the setting and move to the next parameter or
press the
key to save the setting and move to the next submenu of the
setting menu or
press the
key to save the setting, quit the device settings, and return to
measurement mode.
(1)
CAL
3
(2)
keys to set the zero point correction:
Setting option
Meaning
oFF
No zero point correction of the
temperature measurement
-5.0 to +5.0°C
Zero point correction of the
temperature measurement in °C
-9.0 to +9.0°F
Zero point correction of the
temperature measurement in °F
6 Device settings
and selecting the submenu
29

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

202710202730

Table of Contents