6.4
Parameterizing the device
6.4.1 Submenu "SEt Corr": offset/zero point correction and slope cor-
rection of the temperature measurement
Overview of the parameters
After calling up the setting menu with
"SEt Corr", use
SAL
Lin
(1)
(2)
Parameter "OFFS": offset/zero point correction of the temperature measurement
Use the
Menu/parameter
Press the
press the
setting menu or
press the
measurement mode.
All manuals and user guides at all-guides.com
to select the parameter to be configured.
mg/l kOhm
mS
µS
CAL
3
nLF
AL Logg
%/K 1/cm
Parameters
"OFFS": offset/zero point correction of the temperature measurement, see
page 29
"SCAL": slope correction of the temperature measurement, see page 30
and
keys to set the zero point correction:
Setting option
oFF
-5.0 to +5.0°C
-9.0 to +9.0°F
key to save the setting and move to the next parameter or
key to save the setting and move to the next submenu of the
key to save the setting, quit the device settings, and return to
6 Device settings
and selecting the submenu
(1)
(2)
Meaning
No zero point correction of the
temperature measurement
Zero point correction of the
temperature measurement in °C
Zero point correction of the
temperature measurement in °F
29