Limpieza Y Almacenamiento - MasterCool 52225-MED Operating Instructions Manual

Multi-function medical/surface infrared thermometer
Table of Contents

Advertisement

0.36°F
f. The "Repeatability" is
4. Measuring surface temperature:
4.1 After power on, press and hold the "ON/MEM" button, and press the "START" button one time for "Infrared thermometer" mode to see
NOTAS IMPORTANTES
display. In this mode, you can get the target surface temperature.
icon stop flashing to be ready for next measurement.
4.2 When you press the "START" button, you will get the real time temperature immediately. If you press and hold the "START" button, the reading of measurement will
be continuously updated.

LIMPIEZA Y ALMACENAMIENTO:

4.3 Applications include temperature measurements for water, milk, cloth, skin or other objects.
Asegúrese de que la sonda esté limpia para garantizar una lectura precisa.
* Note: This mode shows the actual and unadjusted surface temperature which is different from the body temperature.
El lente de la sonda es la parte más delicada del termómetro. Límpielo con mucho cuidado
5. Power off:
-2.5 ~ -3.1°F
(-1.4 ~ -1.7 ° C).
para evitar dañarlo.
5.1 Device will automatically shut off if left idle for more than 1 minute to extend battery life.
Press the "ON/MEM" button to power on the device. Forehead mode is the default mode. You can see the
0.36°F
(0.20 ° C)
5.2 Manually power off the device by pressing the "ON/MEM" button.
and hear two beep sounds (see figure 1). In this mode, you can hold the thermometer within 4 cm from the central forehead (Fig. 2) and press
• Use hisopos de alcohol o algodón humedecidos con 70% a 75% de alcohol para limpiar el lente de la sonda.
Important Notes
the "START" button to get the forehead measurement. The time consuming for measurement might be 1 second. After each forehead
Please make sure the probe is clean to ensure an accurate reading.
• Deje que la sonda se seque completamente por lo menos 1 minuto.
measurement, wait
icon stop flashing to be ready for next measurement.
Press the "ON/MEM" button to power on the device. Forehead mode is the default mode. You can see the
Points for attention:
The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage.
and hear two beep sounds (see figure 1). In this mode, you can hold the thermometer within 4 cm from the central forehead (Fig. 2) and press
• Mantenga la unidad seca y alejada de líquidos y la luz solar directa.
a. Use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens.
a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its "oral-equivalent" value.
the "START" button to get the forehead measurement. The time consuming for measurement might be 1 second. After each forehead
Press the "ON/MEM" button to power on the device. Forehead mode is the default mode. You can see the
b. Before the measurement, the subject should stay in a stable environment for 5 minutes and avoid exercise, bath for 30mins.
b. Allow the probe to fully dry for at least 1 minute.
measurement, wait
icon stop flashing to be ready for next measurement.
Cleaning
• Rango de temperatura de almacenamiento: almacenarlo a temperatura ambiente entre -4~122°F (-20~+50°C),
c. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature.
and hear two beep sounds (see figure 1). In this mode, you can hold the thermometer within 4 cm from the central forehead (Fig. 2) and press
c. Keep the unit dry and away from any liquids and direct sunlight.
Points for attention:
And Storage
d. The "Clinical Bias" is
-2.5 ~ -3.1°F
the "START" button to get the forehead measurement. The time consuming for measurement might be 1 second. After each forehead
d. Storage temperature range: It should be stored at room temperature between -4~122°F (-20~+50°C), RH ≦ 85%
RH≤85% (humedad relativa)
a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its "oral-equivalent" value.
e. The "Limits of Agreement" is 0.98.
measurement, wait
icon stop flashing to be ready for next measurement.
e. The Probe should not be submerged into liquids.
b. Before the measurement, the subject should stay in a stable environment for 5 minutes and avoid exercise, bath for 30mins.
0.36°F
f. The "Repeatability" is
Press the "ON/MEM" button to power on the device. Forehead mode is the default mode. You can see the
• La sonda no debe sumergirse en líquidos.
Points for attention:
c. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature.
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
4. Measuring surface temperature:
and hear two beep sounds (see figure 1). In this mode, you can hold the thermometer within 4 cm from the central forehead (Fig. 2) and press
-2.5 ~ -3.1°F
d. The "Clinical Bias" is
a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its "oral-equivalent" value.
This could make the body temperature measurement lower than usual.
4.1 After power on, press and hold the "ON/MEM" button, and press the "START" button one time for "Infrared thermometer" mode to see
the "START" button to get the forehead measurement. The time consuming for measurement might be 1 second. After each forehead
Mantener el termómetro en la mano por mucho tiempo, puede causar que la sonda de una lectura de temperatura
e. The "Limits of Agreement" is 0.98.
b. Before the measurement, the subject should stay in a stable environment for 5 minutes and avoid exercise, bath for 30mins.
When the "Low Battery" icon indicates the battery is low, the battery should be replaced immediately with AAA *2 pcs batteries.
display. In this mode, you can get the target surface temperature.
c. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature.
0.36°F
f. The "Repeatability" is
measurement, wait
icon stop flashing to be ready for next measurement.
ambiente más alta. Esto podría hacer que la medición de la temperatura corporal sea más baja de lo habitual.
-2.5 ~ -3.1°F
4.2 When you press the "START" button, you will get the real time temperature immediately. If you press and hold the "START" button, the reading of measurement will
d. The "Clinical Bias" is
1. Open the battery cover: Use the thumbs to push battery cover out. (See Figure 1)
4. Measuring surface temperature:
Points for attention:
e. The "Limits of Agreement" is 0.98.
Press the "ON/MEM" button to power on the device. Forehead mode is the default mode. You can see the
be continuously updated.
Please make sure the probe is clean to ensure an accurate reading.
2. Insert the new AAA *2 pcs batteries. (See Figure 2)
4.1 After power on, press and hold the "ON/MEM" button, and press the "START" button one time for "Infrared thermometer" mode to see
a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its "oral-equivalent" value.
Battery replacement
0.36°F
f. The "Repeatability" is
REEMPLAZO DE BATERIAS:
and hear two beep sounds (see figure 1). In this mode, you can hold the thermometer within 4 cm from the central forehead (Fig. 2) and press
4.3 Applications include temperature measurements for water, milk, cloth, skin or other objects.
The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage.
b. Before the measurement, the subject should stay in a stable environment for 5 minutes and avoid exercise, bath for 30mins.
3. Replace the battery cover. (See Figure 3)
display. In this mode, you can get the target surface temperature.
the "START" button to get the forehead measurement. The time consuming for measurement might be 1 second. After each forehead
4. Measuring surface temperature:
* Note: This mode shows the actual and unadjusted surface temperature which is different from the body temperature.
c. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature.
4.2 When you press the "START" button, you will get the real time temperature immediately. If you press and hold the "START" button, the reading of measurement will
a. Use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens.
Press the "ON/MEM" button to power on the device. Forehead mode is the default mode. You can see the
-2.5 ~ -3.1°F
4.1 After power on, press and hold the "ON/MEM" button, and press the "START" button one time for "Infrared thermometer" mode to see
d. The "Clinical Bias" is
measurement, wait
icon stop flashing to be ready for next measurement.
be continuously updated.
b. Allow the probe to fully dry for at least 1 minute.
5. Power off:
and hear two beep sounds (see figure 1). In this mode, you can hold the thermometer within 4 cm from the central forehead (Fig. 2) and press
e. The "Limits of Agreement" is 0.98.
display. In this mode, you can get the target surface temperature.
Points for attention:
4.3 Applications include temperature measurements for water, milk, cloth, skin or other objects.
c. Keep the unit dry and away from any liquids and direct sunlight.
5.1 Device will automatically shut off if left idle for more than 1 minute to extend battery life.
0.36°F
the "START" button to get the forehead measurement. The time consuming for measurement might be 1 second. After each forehead
f. The "Repeatability" is
4.2 When you press the "START" button, you will get the real time temperature immediately. If you press and hold the "START" button, the reading of measurement will
a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its "oral-equivalent" value.
5.2 Manually power off the device by pressing the "ON/MEM" button.
* Note: This mode shows the actual and unadjusted surface temperature which is different from the body temperature.
d. Storage temperature range: It should be stored at room temperature between -4~122°F (-20~+50°C), RH ≦ 85%
Cuando el icono "batería baja" aparece, indica que las baterías están bajas. Deben ser reemplazadas inmediatamente
measurement, wait
be continuously updated.
b. Before the measurement, the subject should stay in a stable environment for 5 minutes and avoid exercise, bath for 30mins.
icon stop flashing to be ready for next measurement.
4. Measuring surface temperature:
Important Notes
e. The Probe should not be submerged into liquids.
c. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature.
5. Power off:
4.3 Applications include temperature measurements for water, milk, cloth, skin or other objects.
Points for attention:
Figure 1
4.1 After power on, press and hold the "ON/MEM" button, and press the "START" button one time for "Infrared thermometer" mode to see
con 2 baterías nuevas AAA.
Please make sure the probe is clean to ensure an accurate reading.
d. The "Clinical Bias" is
-2.5 ~ -3.1°F
5.1 Device will automatically shut off if left idle for more than 1 minute to extend battery life.
a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its "oral-equivalent" value.
display. In this mode, you can get the target surface temperature.
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
* Note: This mode shows the actual and unadjusted surface temperature which is different from the body temperature.
e. The "Limits of Agreement" is 0.98.
Trouble shooting:
The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage.
1. Abra la tapa donde se encuentran las baterías. Use los pulgares para empujar la tapa de la batería. (véase la figura 1)
b. Before the measurement, the subject should stay in a stable environment for 5 minutes and avoid exercise, bath for 30mins.
5.2 Manually power off the device by pressing the "ON/MEM" button.
4.2 When you press the "START" button, you will get the real time temperature immediately. If you press and hold the "START" button, the reading of measurement will
This could make the body temperature measurement lower than usual.
0.36°F
f. The "Repeatability" is
5. Power off:
c. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature.
Important Notes
a. Use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens.
Error Message
be continuously updated.
When the "Low Battery" icon indicates the battery is low, the battery should be replaced immediately with AAA *2 pcs batteries.
-2.5 ~ -3.1°F
5.1 Device will automatically shut off if left idle for more than 1 minute to extend battery life.
d. The "Clinical Bias" is
4. Measuring surface temperature:
2. Remueva las baterías bajas e inserte las 2 baterías nuevas AAA. (véase la figura 2)
b. Allow the probe to fully dry for at least 1 minute.
4.3 Applications include temperature measurements for water, milk, cloth, skin or other objects.
Please make sure the probe is clean to ensure an accurate reading.
Cleaning
Error 5~9, the system is not functioning properly.
e. The "Limits of Agreement" is 0.98.
5.2 Manually power off the device by pressing the "ON/MEM" button.
1. Open the battery cover: Use the thumbs to push battery cover out. (See Figure 1)
4.1 After power on, press and hold the "ON/MEM" button, and press the "START" button one time for "Infrared thermometer" mode to see
c. Keep the unit dry and away from any liquids and direct sunlight.
* Note: This mode shows the actual and unadjusted surface temperature which is different from the body temperature.
The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage.
0.36°F
And Storage
f. The "Repeatability" is
Important Notes
3. Vuelva a colocar la tapa de la batería en su lugar. (véase la figura 3)
display. In this mode, you can get the target surface temperature.
d. Storage temperature range: It should be stored at room temperature between -4~122°F (-20~+50°C), RH ≦ 85%
2. Insert the new AAA *2 pcs batteries. (See Figure 2)
Measurement before device stabilization.
a. Use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens.
5. Power off:
4. Measuring surface temperature:
Please make sure the probe is clean to ensure an accurate reading.
4.2 When you press the "START" button, you will get the real time temperature immediately. If you press and hold the "START" button, the reading of measurement will
e. The Probe should not be submerged into liquids.
b. Allow the probe to fully dry for at least 1 minute.
3. Replace the battery cover. (See Figure 3)
5.1 Device will automatically shut off if left idle for more than 1 minute to extend battery life.
4.1 After power on, press and hold the "ON/MEM" button, and press the "START" button one time for "Infrared thermometer" mode to see
Cleaning
be continuously updated.
The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage.
The ambient temperature is not within the range between 50
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
c. Keep the unit dry and away from any liquids and direct sunlight.
5.2 Manually power off the device by pressing the "ON/MEM" button.
display. In this mode, you can get the target surface temperature.
4.3 Applications include temperature measurements for water, milk, cloth, skin or other objects.
And Storage
a. Use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens.
This could make the body temperature measurement lower than usual.
d. Storage temperature range: It should be stored at room temperature between -4~122°F (-20~+50°C), RH ≦ 85%
Important Notes
4.2 When you press the "START" button, you will get the real time temperature immediately. If you press and hold the "START" button, the reading of measurement will
(1)
In forehead mode: Temperature taken is higher than 108°F (+42.2
b. Allow the probe to fully dry for at least 1 minute.
* Note: This mode shows the actual and unadjusted surface temperature which is different from the body temperature.
e. The Probe should not be submerged into liquids.
Cleaning
When the "Low Battery" icon indicates the battery is low, the battery should be replaced immediately with AAA *2 pcs batteries.
(2)
In surface mode: Temperature taken is higher than 176°F (+80
be continuously updated.
Please make sure the probe is clean to ensure an accurate reading.
c. Keep the unit dry and away from any liquids and direct sunlight.
5. Power off:
And Storage
1. Open the battery cover: Use the thumbs to push battery cover out. (See Figure 1)
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
4.3 Applications include temperature measurements for water, milk, cloth, skin or other objects.
(1)
In forehead mode: Temperature taken is lower than 93.2°F (+34
d. Storage temperature range: It should be stored at room temperature between -4~122°F (-20~+50°C), RH ≦ 85%
The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage.
5.1 Device will automatically shut off if left idle for more than 1 minute to extend battery life.
(2)
This could make the body temperature measurement lower than usual.
In surface mode: Temperature taken is lower than –7.6°F (-22
2. Insert the new AAA *2 pcs batteries. (See Figure 2)
* Note: This mode shows the actual and unadjusted surface temperature which is different from the body temperature.
e. The Probe should not be submerged into liquids.
a. Use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens.
Battery replacement
5.2 Manually power off the device by pressing the "ON/MEM" button.
Figure 1
When the "Low Battery" icon indicates the battery is low, the battery should be replaced immediately with AAA *2 pcs batteries.
3. Replace the battery cover. (See Figure 3)
b. Allow the probe to fully dry for at least 1 minute.
5. Power off:
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
Important Notes
Device cannot be powered on to the ready stage.
Cleaning
c. Keep the unit dry and away from any liquids and direct sunlight.
1. Open the battery cover: Use the thumbs to push battery cover out. (See Figure 1)
This could make the body temperature measurement lower than usual.
5.1 Device will automatically shut off if left idle for more than 1 minute to extend battery life.
Please make sure the probe is clean to ensure an accurate reading.
And Storage
d. Storage temperature range: It should be stored at room temperature between -4~122°F (-20~+50°C), RH ≦ 85%
2. Insert the new AAA *2 pcs batteries. (See Figure 2)
5.2 Manually power off the device by pressing the "ON/MEM" button.
When the "Low Battery" icon indicates the battery is low, the battery should be replaced immediately with AAA *2 pcs batteries.
Battery replacement
The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage.
Warranty:
e. The Probe should not be submerged into liquids.
Important Notes
TROUBLESHOOTING:
3. Replace the battery cover. (See Figure 3)
1. Open the battery cover: Use the thumbs to push battery cover out. (See Figure 1)
a. Use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens.
Warranty: 12 months
Error 5~9, the system is not functioning properly.
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
Please make sure the probe is clean to ensure an accurate reading.
b. Allow the probe to fully dry for at least 1 minute.
2. Insert the new AAA *2 pcs batteries. (See Figure 2)
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)
Cleaning
Battery replacement
This could make the body temperature measurement lower than usual.
Error Message
Problem
The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage.
Figure 1
c. Keep the unit dry and away from any liquids and direct sunlight.
Ex.SN:E512A000001, the first "E" is External, the second number "5" is the last number of manufacture year, the third and the fourth number "12" is the manufacture
3. Replace the battery cover. (See Figure 3)
And Storage
Measurement before device stabilization.
When the "Low Battery" icon indicates the battery is low, the battery should be replaced immediately with AAA *2 pcs batteries.
a. Use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens.
d. Storage temperature range: It should be stored at room temperature between -4~122°F (-20~+50°C), RH ≦ 85%
month, the others is the serial number.
Trouble shooting:
b. Allow the probe to fully dry for at least 1 minute.
Error 5-9 el sistema no esta fundionando bien
1. Open the battery cover: Use the thumbs to push battery cover out. (See Figure 1)
Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected
e. The Probe should not be submerged into liquids.
Cleaning
The ambient temperature is not within the range between 50
Error Message
c. Keep the unit dry and away from any liquids and direct sunlight.
events at any time, please contact the dealers or nearest service address.
2. Insert the new AAA *2 pcs batteries. (See Figure 2)
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
Figure 1
And Storage
Battery replacement
d. Storage temperature range: It should be stored at room temperature between -4~122°F (-20~+50°C), RH ≦ 85%
Warning: No modification of this equipment is allowed.
(1)
In forehead mode: Temperature taken is higher than 108°F (+42.2
Error 5~9, the system is not functioning properly.
This could make the body temperature measurement lower than usual.
3. Replace the battery cover. (See Figure 3)
e. The Probe should not be submerged into liquids.
Trouble shooting:
(2)
In surface mode: Temperature taken is higher than 176°F (+80
When the "Low Battery" icon indicates the battery is low, the battery should be replaced immediately with AAA *2 pcs batteries.
Please read the instructions for use
Figure 1
Measurement before device stabilization.
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
Error Message
(1)
In forehead mode: Temperature taken is lower than 93.2°F (+34
1. Open the battery cover: Use the thumbs to push battery cover out. (See Figure 1)
Radiant Innovation Inc.
This could make the body temperature measurement lower than usual.
(2)
In surface mode: Temperature taken is lower than –7.6°F (-22
Trouble shooting:
Error 5~9, the system is not functioning properly.
2. Insert the new AAA *2 pcs batteries. (See Figure 2)
Http://www.radiantek.com.tw
The ambient temperature is not within the range between 50
Battery replacement
When the "Low Battery" icon indicates the battery is low, the battery should be replaced immediately with AAA *2 pcs batteries.
Mide temperatura antes que el termómetro se
Error Message
3. Replace the battery cover. (See Figure 3)
Add: 1F, No.3, Industrial East 9
Road,
th
Device cannot be powered on to the ready stage.
Measurement before device stabilization.
(1)
In forehead mode: Temperature taken is higher than 108°F (+42.2
1. Open the battery cover: Use the thumbs to push battery cover out. (See Figure 1)
Figure 1
Science-Based Industrial Park, HsinChu, Taiwan 300.
estabilize
Error 5~9, the system is not functioning properly.
(2)
In surface mode: Temperature taken is higher than 176°F (+80
2. Insert the new AAA *2 pcs batteries. (See Figure 2)
Battery replacement
The ambient temperature is not within the range between 50
(1)
In forehead mode: Temperature taken is lower than 93.2°F (+34
Trouble shooting:
3. Replace the battery cover. (See Figure 3)
Measurement before device stabilization.
La temperatura ambiente no esta dentro del
(2)
In surface mode: Temperature taken is lower than –7.6°F (-22
Error Message
(1)
In forehead mode: Temperature taken is higher than 108°F (+42.2
(2)
rango de 50˚F~104˚F (10˚C~40˚C).
In surface mode: Temperature taken is higher than 176°F (+80
The ambient temperature is not within the range between 50
Device cannot be powered on to the ready stage.
Error 5~9, the system is not functioning properly.
Figure 1
(1)
In forehead mode: Temperature taken is lower than 93.2°F (+34
(1)
In forehead mode: Temperature taken is higher than 108°F (+42.2
(2)
In surface mode: Temperature taken is lower than –7.6°F (-22
Measurement before device stabilization.
Trouble shooting:
(2)
In surface mode: Temperature taken is higher than 176°F (+80
Warranty:
(1)
Figure 1
In forehead mode: Temperature taken is lower than 93.2°F (+34
Error Message
(1) Modo de Frente (corporal): la temperatura
Warranty: 12 months
Device cannot be powered on to the ready stage.
The ambient temperature is not within the range between 50
(2)
In surface mode: Temperature taken is lower than –7.6°F (-22
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)
registrada es mas alta que 108˚F (42.2˚C)
Error 5~9, the system is not functioning properly.
Trouble shooting:
(1)
In forehead mode: Temperature taken is higher than 108°F (+42.2
Ex.SN:E512A000001, the first "E" is External, the second number "5" is the last number of manufacture year, the third and the fourth number "12" is the manufacture
Warranty:
Device cannot be powered on to the ready stage.
Error Message
(2)
(2) Modo: Superficie la temperatura registrada es
In surface mode: Temperature taken is higher than 176°F (+80
month, the others is the serial number.
Measurement before device stabilization.
Warranty: 12 months
Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected
(1)
In forehead mode: Temperature taken is lower than 93.2°F (+34
Error 5~9, the system is not functioning properly.
mas alta que 176˚F (80˚C)
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)
(2)
In surface mode: Temperature taken is lower than –7.6°F (-22
events at any time, please contact the dealers or nearest service address.
Warranty:
The ambient temperature is not within the range between 50
Ex.SN:E512A000001, the first "E" is External, the second number "5" is the last number of manufacture year, the third and the fourth number "12" is the manufacture
Warranty: 12 months
Warning: No modification of this equipment is allowed.
Measurement before device stabilization.
month, the others is the serial number.
(1)
(1) Modo de Frente (corporal): la temperatura
In forehead mode: Temperature taken is higher than 108°F (+42.2
Device cannot be powered on to the ready stage.
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)
Road,
th
Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected
Please read the instructions for use
(2)
In surface mode: Temperature taken is higher than 176°F (+80
Ex.SN:E512A000001, the first "E" is External, the second number "5" is the last number of manufacture year, the third and the fourth number "12" is the manufacture
The ambient temperature is not within the range between 50
registrada es mas baja que 93.2˚F (+34˚C)
events at any time, please contact the dealers or nearest service address.
(1)
In forehead mode: Temperature taken is lower than 93.2°F (+34
Radiant Innovation Inc.
Warranty:
month, the others is the serial number.
Warning: No modification of this equipment is allowed.
(1)
In forehead mode: Temperature taken is higher than 108°F (+42.2
(2)
In surface mode: Temperature taken is lower than –7.6°F (-22
(2) Modo: Superficie la temperatura registrada es
Http://www.radiantek.com.tw
Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected
Warranty: 12 months
(2)
In surface mode: Temperature taken is higher than 176°F (+80
Add: 1F, No.3, Industrial East 9
Road,
Please read the instructions for use
events at any time, please contact the dealers or nearest service address.
th
mas baja que -7.6˚F (-22˚C)
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)
Device cannot be powered on to the ready stage.
(1)
In forehead mode: Temperature taken is lower than 93.2°F (+34
Science-Based Industrial Park, HsinChu, Taiwan 300.
Ex.SN:E512A000001, the first "E" is External, the second number "5" is the last number of manufacture year, the third and the fourth number "12" is the manufacture
Warning: No modification of this equipment is allowed.
Radiant Innovation Inc.
(2)
In surface mode: Temperature taken is lower than –7.6°F (-22
month, the others is the serial number.
Http://www.radiantek.com.tw
Termómetro no se puede encender para su
Please read the instructions for use
Warranty:
Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected
Device cannot be powered on to the ready stage.
Add: 1F, No.3, Industrial East 9
Road,
th
Warranty: 12 months
funcionamiento correcto
Radiant Innovation Inc.
events at any time, please contact the dealers or nearest service address.
Science-Based Industrial Park, HsinChu, Taiwan 300.
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)
Http://www.radiantek.com.tw
Warning: No modification of this equipment is allowed.
Warranty:
Ex.SN:E512A000001, the first "E" is External, the second number "5" is the last number of manufacture year, the third and the fourth number "12" is the manufacture
Add: 1F, No.3, Industrial East 9
Road,
th
Please read the instructions for use
Warranty: 12 months
month, the others is the serial number.
Science-Based Industrial Park, HsinChu, Taiwan 300.
DECLARACION DE EMISIONES ELECTROMAGNETICAS DEL FABRICANTE
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)
Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected
Radiant Innovation Inc.
Ex.SN:E512A000001, the first "E" is External, the second number "5" is the last number of manufacture year, the third and the fourth number "12" is the manufacture
events at any time, please contact the dealers or nearest service address.
Para uso en ambiente electromagnetico (para cuidados de salud en el hogar) especificado mas abajo. El cliente o usuario
Http://www.radiantek.com.tw
month, the others is the serial number.
Warning: No modification of this equipment is allowed.
Add: 1F, No.3, Industrial East 9
Road,
th
debe asegurase que es usado en el ambiente adecuado.
Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected
Science-Based Industrial Park, HsinChu, Taiwan 300.
Please read the instructions for use
events at any time, please contact the dealers or nearest service address.
Radiant Innovation Inc.
Warning: No modification of this equipment is allowed.
Http://www.radiantek.com.tw
Please read the instructions for use
(0.20 ° C)
(-1.4 ~ -1.7 ° C).
(0.20 ° C)
(-1.4 ~ -1.7 ° C).
(0.20 ° C)
(-1.4 ~ -1.7 ° C).
(0.20 ° C)
(-1.4 ~ -1.7 ° C).
(0.20 ° C)
Figure 2
(-1.4 ~ -1.7 ° C).
(0.20 ° C)
Problem
(-1.4 ~ -1.7 ° C).
(0.20 ° C)
Figure 2
Problem
Figure 2
o
F~104
Problem
Figure 2
BF type applied part
Figure 2
Problem
Problem
Figure 2
Ref No.:032020
Problem
Figure 2
Figure 2
Problem
Problem
BF type applied part
BF type applied part
Ref No.:032020
BF type applied part
Ref No.:032020
BF type applied part
Ref No.:032020
BF type applied part
Ref No.:032020
BF type applied part
Ref No.:032020
BF type applied part
icon on the screen
Figure 3
Unload the battery, wait for 1 minute and repower it. If the message
reappears, contact the retailer for service.
Wait for "Er1" to disappear.
Allow the thermometer to rest in a room for at least 15 minutes at room
o
F~104
o
F (10
o
C~40
o
C).
temperature: 50
o
C)
o
C)
Please select the target within specifications. If a malfunction still exists,
please contact the nearest retailer.
o
C)
o
C)
Figure 3
Change with a new battery.
Unload the battery, wait for 1 minute and repower it. If the message
reappears, contact the retailer for service.
Solution
Figure 3
Wait for "Er1" to disappear.
Desconecte las baterias y espere un minuto.
Allow the thermometer to rest in a room for at least 15 minutes at room
o
F (10
o
C~40
o
C).
temperature: 50
Figure 3
Conecte las baterias y encienda el termometro.
Unload the battery, wait for 1 minute and repower it. If the message
o
C)
reappears, contact the retailer for service.
Si el mensaje aparece de nuevo, contacte al
o
C)
Please select the target within specifications. If a malfunction still exists,
Figure 3
Wait for "Er1" to disappear.
please contact the nearest retailer.
o
C)
departamento de servicio.
Unload the battery, wait for 1 minute and repower it. If the message
o
C)
Allow the thermometer to rest in a room for at least 15 minutes at room
o
F~104
o
F (10
o
C~40
o
C).
reappears, contact the retailer for service.
temperature: 50
Espere hasta que desdaparezca el mensaje "Er1"
Change with a new battery.
o
C)
Wait for "Er1" to disappear.
Unload the battery, wait for 1 minute and repower it. If the message
Figure 3
o
C)
Please select the target within specifications. If a malfunction still exists,
reappears, contact the retailer for service.
Allow the thermometer to rest in a room for at least 15 minutes at room
please contact the nearest retailer.
o
F~104
o
F (10
o
C)
o
C~40
o
C).
temperature: 50
Wait for "Er1" to disappear.
Permita que el termómetro se estabilize por 15
o
C)
o
C)
Allow the thermometer to rest in a room for at least 15 minutes at room
o
C)
Please select the target within specifications. If a malfunction still exists,
minutos a la temperatura ambiente donde se va a
Unload the battery, wait for 1 minute and repower it. If the message
o
F~104
o
F (10
o
C~40
o
C).
Change with a new battery.
temperature: 50
Figure 3
please contact the nearest retailer.
reappears, contact the retailer for service.
o
C)
usar 50°F ~104°F (10˚C~40˚C)
o
C)
o
C)
Wait for "Er1" to disappear.
o
C)
Please select the target within specifications. If a malfunction still exists,
please contact the nearest retailer.
o
Figure 3
C)
Allow the thermometer to rest in a room for at least 15 minutes at room
Change with a new battery.
o
F~104
o
F (10
o
C~40
o
C).
o
C)
temperature: 50
Unload the battery, wait for 1 minute and repower it. If the message
reappears, contact the retailer for service.
o
C)
Change with a new battery.
o
C)
Please select the target within specifications. If a malfunction still exists,
Wait for "Er1" to disappear.
Unload the battery, wait for 1 minute and repower it. If the message
please contact the nearest retailer.
o
C)
Por favor seleccione lo requerido dentro de las
Allow the thermometer to rest in a room for at least 15 minutes at room
reappears, contact the retailer for service.
o
C)
o
F~104
o
F (10
o
C~40
o
C).
especificaciones. Si el problema continua, por
temperature: 50
Wait for "Er1" to disappear.
o
C)
Change with a new battery.
favor contacte el departamento de servicio.
Allow the thermometer to rest in a room for at least 15 minutes at room
o
C)
Please select the target within specifications. If a malfunction still exists,
o
F~104
o
F (10
o
C~40
o
C).
temperature: 50
please contact the nearest retailer.
o
C)
o
C)
o
C)
o
C)
Please select the target within specifications. If a malfunction still exists,
Change with a new battery.
please contact the nearest retailer.
o
C)
o
C)
Cambie las baterias
Change with a new battery.
www.mastercool.com
icon on the screen
icon on your LCD
icon on the screen
icon on the screen
icon on the screen
icon on the screen
icon on the screen
Solution
o
F~104
o
F (10
o
C~40
o
C).
Solution
Solution
o
F~104
o
F (10
o
C~40
o
C).
Solution
o
F~104
o
F (10
o
C~40
o
C).
Solution
o
F~104
o
F (10
o
C~40
o
C).
Solution
o
F~104
o
F (10
o
C~40
o
C).
Solution
o
F~104
o
F (10
o
C~40
o
C).
Solution
o
F~104
o
F (10
o
C~40
o
C).
o
F~104
o
F (10
o
C~40
o
C).
icon on your LCD
Figure 2
Figure 2
Figure 2
Figure 2
icon on your LCD
Figure 2
icon on your LCD
Figure 2
icon on your LCD
Figure 2
icon on your LCD
icon on your LCD
icon on your LCD
19