Electrical Connections; Power Cord - Agilent Technologies IDP-3 User Manual

Dry scroll vacuum pump
Table of Contents

Advertisement

Technical Information
There are four versions of the IDP-3:
100 VAC, 50/60 Hz
115 VAC, 60 Hz
220-230 VAC, 50/60 Hz
24 VDC.
Ensure that the pump version conforms to the supply voltage before using.
CAUTION

Electrical Connections

Power Cord

Several power cord options are available from your Agilent dealer. Descriptions of the
available power cords and their ordering numbers are given in the following table.
For high voltage operation, the pump must be connected to the power supply using
NOTE
a high voltage IEC-320 type power cord of at least 10 A capacity.
Be certain that your electrical mains power voltage corresponds to that
CAUTION
indicated on the rear of the pump.
Table 2
Country
Europe
Denmark
Switzerland
UK/Ireland
India
Israel
Japan
North America
160
IDP-3 Power Cord Selection
Power Cord Specification
10 A / 220-230 VAC, 2.5 m
10 A / 220-230 VAC, 2.5 m
10 A / 230 VAC, 2.5 m
13 A / 230 VAC, 2.5 m
10 A / 220-250 VAC, 2.5 m
10 A / 230 VAC, 2.5 m
12 A / 100 VAC, 2.3 m
15 A / 125 VAC, 2.0 m
10 A / 230 VAC, 2.5 m