Optimizing the Stack Configuration

If your module is part of a complete Agilent 1290 Infinity Liquid
Chromatograph, you can ensure optimum performance by installing the
following configurations. These configurations optimize the system flow path,
ensuring minimum delay volume.
For other possible configurations, please refer to the Agilent 1290 Infinity
System Manual.

One Stack Configuration

Ensure optimum performance by installing the modules of the Agilent
1290 Infinity Quaternary LC System in the following configuration (see
Figure 7
flow path for minimum delay volume and minimizes the bench space required.
The Agilent 1290 Infinity Quaternary Pump should always be installed at the
bottom of the stack.
1290 Infinity Quaternary Pump User Manual
on page 32 and
Figure 8

Optimizing the Stack Configuration

on page 33). This configuration optimizes the
Installing the Module
3
31