7
Error Information
General Error Messages

Leak

Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted
temperature-compensation sensor) are used by the leak algorithm to determine
whether a leak is present. When a leak occurs, the leak sensor is cooled by the
solvent. This changes the resistance of the leak sensor which is sensed by the
leak sensor circuit on the main board.
Probable cause
1
2
3
4

Open Cover

Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam
is removed, the fan is switched off, and the error message is generated.
Probable cause
1
2
3
4
1290 Infinity II High-Speed Pumps User Manual
Loose fittings.
Broken capillary.
Loose or leaking purge valve, inlet valve, or
outlet valve.
Defective pump seals.
The top foam was removed during operation. Please contact your Agilent service representa-
Foam not activating the sensor.
Defective sensor or main board.
Rear of the module is exposed to strong
direct sunlight.
Suggested actions
Ensure all fittings are tight.
Exchange defective capillaries.
Ensure pump components are seated correctly.
If there are still signs of a leak, exchange the
appropriate seal (purge valve, inlet valve, outlet
valve).
Exchange the pump seals.
Suggested actions
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Ensure that the rear of module is not directly
exposed to strong sunlight.