Replacing the Progard Cartridge and Vent Filter,
Follow the steps below to install a new Progard Cartridge.
Placing
Step
Action
1
Remove the covers on the 2
ports of the Progard Cartridge.
Wet the O-rings with water.
2
Install the Progard Cartridge
until it is fully seated.
Close the right door.
3
When a new Progard Cartridge
is installed, the LCD looks like
this.
4
Press
.
5
When the Progard Cartridge
flush has finished, the Water
System goes to READY Mode.
- 76 -