Specifications and requirements,
The Feed water requirements are listed here.
Feed water
The Environmental requirements are listed here.
Environmental
Ambient operating temperature
Relative humidity during storage
The noise level is < 50 dB at a distance of 1 metre.
Noise Level
The minimum consumables required for installation are listed here.
Consumables
Note that these items are not shipped with the System and must be ordered
separately:
Progard Cartridge,
Q-Pak Pack, and
POD Pak.
The Reservoir must be located relative to the Water System:
Reservoir
0 y 2 metres, where y = vertical distance, and
location
0 x 3 metres, where x = horizontal distance.
Parameter
Type of Feed water
Conductivity
Pressure
Temperature
Dissolved CO
Free Chlorine
Fouling Index
pH
Parameter
Altitude
Ambient storage temperature
Installation Category
Location
Pollution Degree
and operation
Continued
2
The System is intended for indoor use
Maximum relative humidity 80% for
temperatures up to 31°C decreasing
linearly to 50% relative humidity at
- 18 -
Value
Potable tap water
< 2000 S/cm
1 bar < P < 6 bar
5°C < T < 35°C
< 30 ppm
< 3 ppm
< 12
4 < pH < 10
Value
< 3000 metres
4 – 40°C
4 – 40°C
II
only.
2
40°C.