Millipore Milli-Q Direct 8 System User Manual page 43

Table of Contents

Advertisement

Calibrating the Flow rate,
Procedure
(continued)
Step
10
11
Continued
Action
6
Place a 1 L Graduated Cylinder
under the P
OD Unit.
Press
.
7
Press
.
8
Water
dis nses automatically
pe
from the POD Unit.
Wait until it stops dispensing
water.
9
Measure the amount of water
(in ml) that was dispensed.
Suppose 8
70 ml was collected.
Input this using the Keypad.
Perfor the flow calibration
m
again to im
prove accu
Press
.
Press 3 times on
racy.
.
- 43 -