Installing the Progard Cartridge

Follow the steps below to install a new Progard Cartridge.
Procedure
Step
Action
1
Start in STANDBY Mode.
NOTE:
The PROGARD CARTRIDGE
OUT Alarm message is not
shown at this time. By
following the instructions
earlier in this manual, the alarm
was cancelled.
2
 Open the right door of the
System Cabinet.
 Remove the 2 protective caps
located on the ports inside.
 Remove the covers on the 2
3
ports of the Progard
Cartridge.
 Wet the O-rings with water.
 Install the Progard Cartridge
4
until it is fully seated.
 Close the right door.
5
One minute later, the Main
LCD shows that a new Progard
Cartridge is installed.
- 26 -