Registering UV Lamp timers

The timer used for each UV Lamp must be
Introduction
If this is not done, then the message indicating th
needed is shown too early.
The UV Lamp timers need to be reset for:
 the UV 185 nm Lamp, and
 the UV 254 nm Lamp.
NOTE:
Before doing this, make sure that the date and tim
accuracy.
This procedure shows how to reset the tim
Procedure
Step
Action
1
Place the System in STANDBY
Mode.
2
Select Menu.
Press
.
3
Select Maintenance.
Press
.
4
Select Install UV 185 nm
Lamp.
Press
.
reset when the System
at a Lamp replacem
er used for the UV 185 nm
- 36 -
is installed.
ent is
e have been checked for
Lamp.
Diagram
Continued on next page