Summary of Alarm messages,
Alarm stop
messages (suite)
PROGARD CARTRIDGE OUT
Q-PAK PACK OUT
TANK EMPTY
WATER DETECTED
Suite
LCD message
- 101 -
What it means
 This alarm stops the RO portion of
the system.
 The Progard Cartridge is not
installed correctly or it has been
removed.
 Verify that the Progard Cartridge
is installed correctly.
 Contact Millipore if the problem
continues.
 This alarm stops the Milli-Q
p
ortion o
f the system.
 The Q-Pak Pack is not installed
correctly or it has been removed.
 Verify that the Q-Pak Pack is
installed correctly.
 Contact Millipore if the problem
continues.
 This alarm stops the Milli-Q
portion of the system.
 The System has detected an empty
Reservoir.
 Refill the Reservoir.
 Verify that the Reservoir level
sensor is plugged into the System
Cabinet.
 This alarm stops the whole system
to prevent risks of flood.
 A Water Sensor (an accessory
connected to the System) has
detected water on the surface
where it is located. This may be
caused by a leak.
 Clean up the spilled water.
 Make sure the source of the leak is
fixed.
 Place the system in Standby
Mode, then Ready mode.