Rinsing the Q-Pak Pack

There has to be enough water in the Reservoir in order to rinse the Q-Pak
Pack.
Have enough
water!
If there is not enough water in the Reservoir, the TANK EMPTY Alarm is
shown. Additionally, air can enter the tubings and can temporarily affect
other sensors.
Follow the steps below to rinse the Q-Pak Pack.
Procedure
Step
Millipore Reservoir
30 Litre
60 Litre
100 Litre
Action
 Locate the clear tubing and
1
the barbed fitting from the
System Accessories Bag.
 Screw the barbed fitting onto
the POD Unit.
 Push one end of the clear
tubing onto the end of the
barbed fitting.
 Place the other end of the
clear tubing into a sink.
NOTE:
Do not use any white tape on
the threads of the barbed fitting.
An O-ring located inside the
POD Dispenser ensures water
tightness.
2
Place the System into READY
Mode.
3
Push the POD Plunger all the
way down and then release it.
In a few minutes, water should
come out of the POD Unit.
> 40% Tank Level
> 30% Tank Level
- 32 -
Minimum
100% Tank Level
Diagram
Continued on next page