Summary of Alert messages,
Alert messages
(continued)
PERFORM TOC CURVE
CHECK
QUALIFICATION VISIT
OVERDUE XX DAYS
REPLACE ASM UV LAMP
REPLACE ASM UV LAMP IN
XX DAYS
REPLACE EXTERNAL
PRETREATMENT
REPLACE EXTERNAL
PRETREATMENT IN XX
DAYS
REPLACE POD PAK
REPLACE POD PAK IN XX
DAYS
Continued
LCD message
- 110 -
What it means
 The System has determined that a
TO
C Curv
e Check should
performed.
Perform a TOC Curve Check or wait
until one is done automatically.
 If this message persists after 24
hours of its appearance, then contact
Millipore.
 The System has determined that a
Qualification Visit is overdue.
 Contact Millipore.
 The Water System has determined
that the ASM UV Lamp should be
rep
laced.
 Contact Millipore.
The Water System has determined
that the ASM UV Lamp on the
Reservoir should be replaced in XX
days, where XX is 15, ..., 1.
 Contact Millipore.
 The Water System has determined
that the external pretreatment
(optional) should be replaced.
 Consult the documentation supplied
with the external pretreatment for
more information.
 The Water System has determined
that the external pretreatment
(optional) should be replaced in XX
days, where XX is 15, ..., 1.
 Consult the documentation supplied
with the external pretreatment for
more information.
 The System has determined that the
POD PAK needs replacement.
 Replace the POD Pak and reset the
timer.
 The System has determined that the
POD PAK should be replaced in XX
days, where XX is 15, ..., 1.
 Replace the POD Pak and reset the
timer.
Continued on next page
be