Summary of Alarm messages
Alarm stop
messages
FLOW AUTO STOP
INCORRECT PROGARD
CARTRIDGE
INCORRECT Q-PAK PACK
LOW FEED WATER PRESSURE
PERMEATE C > SP
POD LOCKED
LCD message
- 100 -
What it means
This alarm stops the Milli-Q
portion of the system.
A safety feature of the Milli-Q
Direct system has automatically
stopped dispensing
minutes to avoid risks of overflow.
P
ush the
POD Unit Plung
way down and release.
This resets the dispenser timer and
makes the POD Unit available for
dispensing.
This alarm stops the RO portion of
the system.
The System does not recognize the
type of Progard Cartridge being
installed.
Contact Mi
llipore.
This alarm stops the Milli-Q
portion of the
system.
The System does not recognize the
type of Q-Pak Pack being
installed.
Contact Millipore.
This alarm stops the RO portion of
the system.
Check Feed water pressure and
rectify.
Go to STANDBY Mode and go to
READY Mode to release any
trapped air in the Water System.
Contact Millipore if the problem
persists.
This alarm stops the RO portion of
the system
.
Th
e Permeate conductivity is
ab
ove the set point.
Co
ntact Millipore.
This alarm
stops the Milli-Q
portion of the sys
tem.
The POD Unit was left in the open
position.
Push the Plunger all the way down
and release.
Continued on next page
water after 40
er all the