Specifications and requirements
The water delivered from a POD Unit has the following characteristics.
Milli-Q
Water
quality
Particulates > 0.22 m**
(*) With BioPak
(**) With Millipak
NOTE:
These specifications are valid if feed water within specification and if correct
maintenance is performed on the system. Some specifications may not be
achieved at start-up.
The various weights are found in the table below.
Weight
The electrical specifications and data are found in the table below.
Electrical
Power Cord Length
Electrical Ground
Parameter
Resistivity
TOC
Bacteria**
Pyrogens*
RNases*
DNases*
Flow Rate**
Final Filter
or BioPak Final Filter
System
Milli-Q Direct 8
Milli-Q Direct 16
Parameter
Voltage
Frequency
3.15 Amp Fast Acting; 5 mm x 20 mm; 250 V
Main Fuse
The fuse should be serviced by a qualified
Power Used
The System is powered on and off by removing the
Power Cord use
The power cord should be plugged into a wall
Specification
18.2
5
< 1
< 0.1
< 0.001
< 0.01
< 4
0.05 – 1.5
Operating
Dry Weight
Weight (kg)
27
28
100-230 VAC 10%
50-60 Hz 10%
safety voltage.
Millipore Service Representative.
145 VA
2.5 metres
Earth Grounded
power cord from the wall outlet.
outlet that is accessible.
- 16 -
Units
M.cm @25°C
ppb
Particulates/mL
cfu/mL
Eu/mL
ng/mL
pg/L
L/min
Shipping
(kg)
Weight (kg)
20
24
21
25
Value
Continued on next page