Alarms generated during installation
During the installation of a Milli-Q System, certain Alarm messages are
Overview
generated.
This occurs because:
These alarms are explained here. For more information about Alarm
messages, see the chapter titled 'Alarms'.
This alarm occurs because the Reservoir is empty during most of the
TANK EMPTY
installation.
message
This alarm goes away when the Reservoir is partially full.
To cancel the text display of this alarm message, follow the instructions on
the LCD.
This alarm occurs because the Progard Cartridge is not installed.
PROGARD
This alarm goes away when the Progard Cartridge is detected by the Milli-Q
CARTRIDGE
OUT message
System.
To cancel the text display of this alarm message, follow the instructions on
the LCD.
This alarm occurs because the Q-Pak Pack is not installed.
Q-PAK PACK
This alarm goes away when the Q-Pak Pack is detected by the System.
OUT message
To cancel the text display of this alarm message, follow the instructions on
the LCD.
This alarm occurs because the Q-Pak Pack is not fully rinsed out or there is
MILLI-Q RES
air in the tubing near the resistivity sensor.
< SP,
REPLACE
This alarm goes away when a few litres of water are dispensed from the
Q-PAK
POD Unit.
message
To cancel the text display of this alarm message, follow the instructions on
the LCD.
This alarm occurs because there is air in the tubings and in the new Progard
LOW FEED
cartridge.
WATER
PRESSURE
When the air is gone and replaced with water, this alarm does not occur
message
anymore during installation.
To cancel the text display of this alarm message, follow the instructions on
the LCD.
o the Reservoir is empty,
o there is air in the tubings and in the Progard Cartridge,
o the Progard Cartridge is not installed, and
o the Q-Pak Pack is not installed.
- 20 -
Continued on next page