Download Print this page

Optional Dmc Ozonator; Dmc Ozonator Removal - Thermo Scientific 49iQ Instruction Manual

Ozone analyzer

Advertisement

Optional DMC
Ozonator
DMC Ozonator
Thermo Scientific
Use the following procedure to remove the DMC from the instrument
case. It is easier to do the following by removing the DMC first then
remove and replace the following as necessary:
Ozonator DMC removal
Lamp replacement
Ozonator DMC board replacement
Ozonator detector board replacement
Ozonator replacement
Ozonator DMC reassemble
Use the following procedure to remove and replace the ozonator.
Removal
Equipment required:
Phillips drive, #2
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2. Unplug electrical, unplug DMC cable to ozonator board DMC (J4).
3. Disconnect 2X plumbing.
4. Using a #2 Phillips drive, unfasten captive hardware.
2–1).