Detector Board Replacement; Figure 7-42. Replacing Detector Board - Thermo Scientific 49iQ Instruction Manual

Ozone analyzer

Advertisement

Servicing
49iQ Photometer DMC
Detector Board
Replacement
7-36
49iQ Instruction Manual
Use the following procedure to replace the detector board.
Equipment required:
Hex drive, 3/32
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2–1).
2. Unplug electrical cable from detector 1 and detector 2.
3. Using a 3/32 hex drive, unfasten four #4-40 cap head screws to remove
the cover.
4. Remove detector cover.
5. Using a 3/32 hex drive, unfasten four #4-40 cap head screws to replace
per board.
Figure 7–42. Replacing Detector Board
6. Replace detector board and reassemble in reverse order.
Thermo Scientific

Advertisement

loading

This manual is also suitable for:

Iq series