Download Print this page

Photometer Dmc Board Replacement; Figure 7-39. Photometer Board Replacement - Thermo Scientific 49iQ Instruction Manual

Ozone analyzer

Advertisement

Servicing
49iQ Photometer DMC
Photometer DMC
Use the following procedure to replace the board.
Board Replacement
Equipment required:
Phillips drive, #2
1. Unplug electrical, detector 1, detector 2, thermistor, and heater.
2. Using a #2 Phillips drive, unfasten five 6-32 pan head screws.
Figure 7–39. Photometer Board Replacement
3. Replace board and reassemble in reverse order.
Thermo Scientific
49iQ Instruction Manual
7-33