Thermo Scientific 49iQ Instruction Manual page 163

Ozone analyzer

Advertisement

Thermo Scientific
4. Allow this value to stabilize.
5. Record the value as [O
2 in the
"Periodic Zero and Span
Note Level 1 setting.
6. Adjust Level 2 for the desired level (typically 90 ppb).
7. Allow this value to stabilize.
8. Record the Level 2 value as [O
equation:
URL
[(A
Z)
20
%
Error
[O
Where:
A
= Recorder response with precision level, % scale
20
Z = Recorder response with zero air, % scale
URL = Upper range limit
Note The expected stability of the instrument section is greater then the
expected stability of the internal ozonator.
]
to be used in the span error equation of Step
3
80
Checks" section described previously.
]
to be used in the following error
3
20
]
[O
]
3
20
100
100
]
3
20
Calibration
Internal Ozonator Adjustment
49iQ Instruction Manual
4-15