Operation
Settings
high range is active, the range mode selection status output is at half of
full-scale.
When the high range is active, the concentration must drop to 85% of
the low O
range for the low range to become active.
3
In addition, each O
range and analog output has a span coefficient so
3
that each range can be calibrated separately. This is especially necessary
if the two ranges are not close to one another. For example, the low O
3
range is set to 0–50 ppb and the high O
range is set to 0–20,000 ppb.
3
Thermo Scientific
49iQ Instruction Manual
3-73