Figure 7-31. Clear Partition Keyway Step Pol Board - Thermo Scientific 49iQ Instruction Manual

Ozone analyzer

Advertisement

Thermo Scientific
Figure 7–31. Clear Partition Keyway Step POL Board
6. Slide step board 1 upwards clearing the partition panel keyway.
7. If replacing step board 1, make sure switch 1 and 2 are pointed away
from ON
(Figure
7–32). If replacing optional step board 2, make sure
switch 1 is pointed towards ON and switch 2 is pointed away from ON
(Figure
7–33).
8. Replace Step POL board and reassemble in reverse order.
Servicing
Step POL Board Replacement
49iQ Instruction Manual
7-27