Alarm Setpoints - Thermo Scientific 49iQ Instruction Manual

Ozone analyzer

Advertisement

Operation
Settings

Alarm Setpoints

3-96
49iQ Instruction Manual
The Alarm Setpoints screen allows the user to view and set all settable
alarm minimum and maximum values. Alarm setpoints can also be set in
Settings>Health Check>Status and Alarms screens.
Note User cannot set alarm limits outside of the acceptable range. The
minimum and maximum alarm limit can also be set by pressing on the
corresponding buttons located in the Settings>Health Check>Status and
Alarms screen. See
"Status and
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Instrument Settings>Alarm Setpoints
The Alarm Setpoints screen contains the following information:
Across:
Variable: Lists the items that have settable alarm limits.
Value: Displays the current value for each item.
Low Alarm: User sets low alarm for item.
High Alarm: User sets high alarm for item.
Units: Units for each item (not settable).
Down:
O
: O
concentration alarm.
3
3
Bkg Check Offset: User can set the maximum allowable background
reading offset for calibration and calibration checks. This is set with
the high alarm only.
Alarms" on page 3-36.