Servicing
Optional DMC Ozonator

Lamp Replacement

7-44
49iQ Instruction Manual
Use the following to replace the lamp.
Equipment required:
Ball-end hex drive, 3/32
1. Remove the ozonator DMC from the instrument. Refer to
Ozonator
Removal" on page 7-41.
2. Using a 3/32 ball-end hex drive (angle alignment of screws), unfasten
two #4-40 cap screws.
3. Pull lamp out.
Figure 7–51. Replacing the Lamp
4. Replace lamp and reassemble in reverse order.
"DMC