13. Trouble Shooting
13.1 Introduction
Error number
0x0203
0x0204
0x0205
0x0206
0x0207
0x0208
0x0209
0x020A
0x020B
0x020C
0x020D
0x020E
0x020F
0x0210
0x0211
0x0212
0x0301
0x0302
2004-10
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
Error text
Possible Defective Part
"Not expected
Optical switch within transport box is
transport FixedOut
broken, plate calibration values got
IRQ"
lost
"Not expected
Optical switch within transport box is
transport Rear
broken, plate calibration values got
IRQ"
lost
"Transport lost %s
Optical switch within transport box is
steps"
broken, plate calibration values got
lost
"Transport
Optical switch within transport box is
inserted %s steps"
broken, plate calibration values got
lost
"Transport can't
Optical switch within transport box is
find HOME
broken, plate calibration values got
position"
lost
"Transport can't
Optical switch within transport box is
find REAR
broken, plate calibration values got
position"
lost
"Lamp high"
too much light to perform a reliable
measurement
"Lamp low"
too less light to perform a reliable
measurement
"Timeout waiting
check power board and lamp
for lamp on"
"Timeout waiting
Problem with the transport system
for measurement
see transport box, optical switches
finished"
broken
"Timeout waiting
Problem with the transport system
for transport
see transport box, optical switches
finished"
broken
"Not expected
Problem with the transport system
transport IRQ %d
see transport box
during calibration"
"Transport couldn't
Problem with the transport system
find full dark edge
see transport box
during calibration"
"Transport lost %d
Problem with the rear sensor of the
steps during
transport box
calibration "
"Transport
check settings of Measurement
frequency too low:
Mode
%d < %d"
"Transport
check settings of Measurement
frequency too
Mode
high: %d > %d"
"Writing to flash
Problem with the main board
failed"
"Verifying flash
Problem with the main board
failed"
13. Trouble Shooting
See chapter...
4.4.6 Plate calibration
4.4.6 Plate calibration
4.4.6 Plate calibration
Duration Test page 4-16
4.4.6 Plate calibration
Duration Test page 4-16
4.4.6 Plate calibration
4.4.6 Plate calibration
4.4.5 Mirror and lamp
adjustment
4.4.5 Mirror and lamp
adjustment
4.4.5 Mirror and lamp
adjustment
7.5 Replacing the
Transport Box
7.5 Replacing the
Transport Box
7.5 Replacing the
Transport Box
7.5 Replacing the
Transport Box
7.5 Replacing the
Transport Box
4.3.5 Define Measurement
Mode
4.3.5 Define Measurement
Mode
8.5Replacing the Main
Board
8.5Replacing the Main
Board
13-1
Need help?
Do you have a question about the sunrise RC and is the answer not in the manual?
Questions and answers