11. Maintenance & Cleaning

11.8 Disinfection Declaration

11-10
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
The following procedure should be used to disinfect the instrument.
1.
Wear protective gloves, protective glasses and protective clothing.
2.
Prepare an autoclaveable bag for all disposables used during the
disinfection procedure and label it with autoclave tape.
3.
Disconnect the instrument from the main power supply to avoid any risk of
explosion.
4.
Disconnect the instrument from the computer.
5.
Unplug the USB stick and close the cover for the USB stick housing.
6.
Carefully spray the disinfectant solution (or use a disposable soft tissue
paper towel soaked in the disinfectant) on all outer surfaces of the
instrument.
7.
After a minimum contact time of 10 minutes, repeat the previous step of this
procedure.
8.
After a contact time of five hours wipe the instrument using a soft paper
towel and a mild detergent or distilled water to remove all traces of the
disinfectant.
9.
Wipe dry the outer surfaces of the instrument.
10. Pack the instrument and its accessories.
11. Disinfect your hands and clean them with a mild detergent.
12. Complete a disinfection declaration and attach it to the outside of the
box so that it is clearly visible. See below for an example of the
disinfection declaration.
See below for an example of the disinfection declaration, which must be
completed before the instrument is returned to the distributor for servicing.
The following label MUST be completed by the operating authority and attached
to the top of the package in which the instrument is returned, before sending the
instrument to the servicing center for repair.
I declare that the instrument in this package has been decontaminated or
disinfected to remove or inactivate any biological material, which could be
dangerous to the service personnel, or that it has never been exposed to
any hazardous biological material.
Name:...................................................................................................................
Firm: .....................................................................................................................
Address: ...............................................................................................................
.............................................................................................................................
.............................................................................................................................
Country:................................................................................................................
Signature:.............................................................................................................
2004-10