10. Sunrise Service Tool Kits for Sunrise RC and Options
10-14
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
Enter the single command "DB MCU POS 24" in the "command" field and press
the send button – Plate transport in BCR position
Enter the single command "DB IBUS SEND B S" in the "command" field and
press the send button – start barcode scanning
Enter a barcode reset (DB IBUS SEND B R) step in the command field
before entering the barcode scanning command.
The barcode scanner will then be switched on for five seconds. During this time,
adjust the laser beam with the mirror screw (see picture below), so that the laser
beam is aligned with the reticule on the adjustment microplate.
The barcode scanner may have to be switched on more than once in order to
align the laser properly.
DB MCU POS 24
Answer: -
Note:
DB IBUS SEND B S
Answer: +
2004-10