4.0 Operation

4.5 Ear irrigation bowl with separate suction channel (530.2070.0)

The ear irrigation bowl with optional, separate suction channel may only be used for suction during ear irrigation or stimulation.
!
Only the ear irrigation bowl with integrated sieve (hole diameter 1 mm at a maximum) may be used. The suction channel is not
suitable for any other suction procedure in the medical field! The liquid pump is not suitable for sucking off bigger particles, for
example, cerumen, etc. Therefore, suction may only be performed with a bowl and sieve.
Greasy and pasty substances can be sucked off but only limited. It is recommended to clean the bowl after each suction
procedure and to rinse it with water (disinfectant) in order to prevent the pump valves from jamming.
When you remove the suction nozzle with the ear irrigation bowl from the suspension, the suction system starts automatically.
Apply the bowl with the lateral cut-out below the ear which needs to be irrigated and start irrigation. When the suction nozzle is
put back to the suspension, the suction system stops automatically.

4.6 Compressed-air system

• Make sure equipment and hoses are firmly connected in order to build up a sufficient pressure.
• If you use politzer olives, control pressure in acc. with the appliance.

4.6.1 Activation

Remove the handle from the support to activate the compressed-air system. The compressed-air system is automatically
activated via the installed light barrier.
22
Ear irrigation module
(compressed air)
Nebulization of medications
(compressed air)
Handle conpressed air